Case Report of Anti-Thrombogenicity

Authors:

Abstract
Background: The Toray Filtryzer™-NF is a new polymethyl methacrylate filter with improved hemocompatibility and reduced impact on platelet counts. Objectives: This suggests that, if necessary, a reduction in anticoagulation may be possible when dialysis is performed with the Toray Filtryzer™-NF. Methods: In the following, we dialyzed 5 hemodialysis patients who had a contraindication to full anticoagulation postoperatively or after renal biopsy with the Filtryzer™-NF. Results: A significant reduction in heparin administration was achieved, and in 1 patient, heparin substitution was completely omitted. Despite the significantly reduced heparin doses, no thrombosis of the system occurred during the hemodialysis. Conclusion: In conclusion, hemodialysis using the Toray Filtryzer™-NF is an effective alternative for patients at significantly increased risk of bleeding.