Radiographic Assessment of Total Knee Arthroplasty Alignment With and Without Accelerometer-Based Navigation at a Resident Training Institution
- 22 November 2024
- journal article
- research article
- Published by Georg Thieme Verlag KG in The Journal of Knee Surgery
- Vol. 38 (05) , 224-228
Abstract
Malalignment and lack of surgeon experience are cited as risk factors for prosthetic failure and the need for subsequent revision surgery in total knee arthroplasty (TKA). Various conventional (CON) and computer-assisted surgical (CAS) methods have been developed to try and prevent malalignment and limit outliers. One of these methods is through an accelerometer-based CAS (aCAS), which intraoperatively helps determine the angulation and amount of resection necessary to restore alignment in TKA. The purpose of this study was to determine the number of TKA alignment outliers in TKAs performed both with CON and aCAS methods at a single training institution in cases that trainees were involved in. In this retrospective single-center study, radiographic analysis of 150 primary TKAs performed at our training institution was performed. The medial distal femoral angle (DFA), medial proximal tibial angle (PTA), and posterior slope angle (PSA) were evaluated on immediate postoperative radiographs of 75 aCAS and 75 CON knees. Outlier measurements were defined as DFA outside of 5 ± 3 degrees valgus, PTA > ± 3 degrees from neutral axis, and PSA outside 0 to 7 degrees for cruciate retaining and 0 to 5 degrees for posterior stabilized implants. Data was analyzed using chi-squared, analysis of variance, and Student's t-tests. There was no significant difference found between these two groups in the total number of outliers (8% aCAS vs. 9.8% CON, p = 0.508). Additionally, there was also no significant difference found for any of the radiographic measurements independently; DFA (8% vs. 8%, p = 1.00), PTA (4% vs. 9%, p = 0.184), and PSA (12% vs. 12%, p = 1.00), when comparing aCAS and CON TKAs. Primary TKAs performed at a single training institution demonstrated no significant difference between aCAS and CON methods in the total number of outliers or in the number of outliers for each postoperative radiographic measurement. This study would suggest that there is no advantage in limiting outliers when utilizing aCAS for TKAs at training institutions in cases that trainees are involved in. Level of Evidence Level III Retrospective Cohort StudyKeywords
This publication has 15 references indexed in Scilit:
- The use of accelerometer-based navigation for coronal TKA alignment: a prospective, single surgeon comparative studyArchives of orthopaedic and trauma surgery, 2019
- Satisfaction Rates and Quality of Life Changes Following Total Knee Arthroplasty in Age-Differentiated CohortsThe Journal of Arthroplasty, 2018
- Accelerometer-Based, Portable Navigation (KneeAlign) vs Conventional Instrumentation for Total Knee Arthroplasty: A Prospective Randomized Comparative TrialThe Journal of Arthroplasty, 2017
- Accuracy of A Handheld Accelerometer-Based Navigation System for Femoral and Tibial Resection in Total Knee ArthroplastyThe Journal of Arthroplasty, 2015
- Extramedullary Guides Versus Portable, Accelerometer-Based Navigation for Tibial Alignment in Total Knee Arthroplasty: A Randomized, Controlled Trial: Winner of the 2013 HAP PAUL AwardThe Journal of Arthroplasty, 2014
- The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesisInternational Orthopaedics, 2013
- Health-Related Quality of Life After Total Joint ArthroplastyClinics in Geriatric Medicine, 2012
- Patient Satisfaction after Total Knee Arthroplasty: Who is Satisfied and Who is Not?Clinical Orthopaedics and Related Research, 2010
- Good Alignment After Total Knee Arthroplasty Leads to Faster Rehabilitation and Better FunctionThe Journal of Arthroplasty, 2009
- Meta-Analysis of Alignment Outcomes in Computer-Assisted Total Knee Arthroplasty SurgeryThe Journal of Arthroplasty, 2007