Clinical and radiographic results and wear performance in different generations of a cementless porous-coated acetabular cup
Authors:
- 31 January 2007
- journal article
- Published by Springer Nature in International Orthopaedics
- Vol. 32 (2) , 181-187
Abstract
We compared clinical results and wear performance in two different generations of a cementless porous-coated cup, analysing the long-term results of 83 uncemented Harris-Galante I cups (32-mm femoral head) and 93 uncemented Harris-Galante II cups (28-mm femoral head). All polyethylene liners were gamma irradiated in air. Polyethylene linear wear was estimated using a software package. The minimum follow-up was 10 years. Nine Harris-Galante I cups and two Harris-Galante II cups were revised due to aseptic loosening or polyethylene problems. The mean femoral head penetration at 6 weeks after surgery was 0.15 ± 0.05 mm for the Harris-Galante I cups and 0.12 ± 0.03 for the Harris-Galante II cups (p < 0.001);but mean wear was 0.13 ± 0.23 mm per year for the Harris-Galante I cups and 0.11 ± 0.10 for the Harris-Galante II cups (p = 0.740). Most of the metallic shells in both groups showed stable fixation. The so-called second-generation cups had lower initial polyethylene wear that resulted in less polyethylene wear at the latest the follow-up, but the overall wear rate was similar in both groups despite the different femoral head sizes and the improved locking mechanism.Keywords
This publication has 25 references indexed in Scilit:
- Precision and Limitation of Measuring Two-dimensional Wear on Clinical RadiographsClinical Orthopaedics and Related Research, 2006
- Long-term results of uncemented acetabular cups with an ACS polyethylene liner. A 14–16-year follow-up studyInternational Orthopaedics, 2006
- Radiographic Signs of Osseointegration in Porous-coated Acetabular ComponentsClinical Orthopaedics and Related Research, 2006
- COMPARISON OF POROUS-COATED TITANIUM FEMORAL STEMS WITH AND WITHOUT HYDROXYAPATITE COATINGJournal of Bone and Joint Surgery, 2003
- The Harris-Galante total hip arthroplastyThe Journal of Arthroplasty, 2001
- A comparison of polyethylene wear in hips with cobalt-chrome or zirconia headsThe Journal of Bone and Joint Surgery. British volume, 2001
- Three-dimensional polyethylene wear of a press-fit titanium prosthesisThe Journal of Arthroplasty, 1997
- Measurement accuracy in acetabular cup wearThe Journal of Arthroplasty, 1995
- Severe polyethylene wear in total hip arthroplastyThe Journal of Arthroplasty, 1994
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958