Ultrasound speckle reduction using modified gabor filters
Authors:
- Ricardo G. DantasRDRicardo G. Dantas
- Eduardo T. CostaECEduardo T. Costa
- 26 March 2007
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 54 (3) , 530-538
Abstract
B-mode ultrasound images are characterized by speckle artifact, which may make the interpretation of images difficult. One widely used method for ultrasound speckle reduction is the split spectrum processing (SSP), but the use of one-dimensional (1-D), narrow-band filters makes the resultant image experience a significant resolution loss. In order to overcome this critical drawback, we propose a novel method for speckle reduction in ultrasound medical imaging, which uses a bank of wideband 2-D directive filters, based on modified Gabor functions. Each filter is applied to the 2-D radio-frequency (RF) data, resulting in a B-mode image filtered in a given direction. The compounding of the filters outputs give rise to a final image in which speckle is reduced and the structure is enhanced. We have denoted this method as directive filtering (DF). Because the proposed filters have effectively the same bandwidth as the original image, it is possible to avoid the resolution loss caused by the use of narrow-band filters, as with SSP. The tests were carried out with both simulated arid real clinical data. Using the signal-to-noise ratio (SNR) to quantify the amount of speckle of the ultrasound images, we have achieved an average SNR enhancement of 2.26 times with simulated data and 1.18 times with real clinical dataKeywords
This publication has 20 references indexed in Scilit:
- Aggressive region growing for speckle reduction in ultrasound imagesPattern Recognition Letters, 2003
- Strain compounding: spatial resolution and performance on human imagesUltrasound in Medicine & Biology, 2001
- A textural approach based on gabor functions for texture edge detection in ultrasound imagesUltrasound in Medicine & Biology, 2001
- In vitro spatial compound scanning for improved visualization of atherosclerosisUltrasound in Medicine & Biology, 2000
- A general statistical model for ultrasonic backscattering from tissuesIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2000
- Using phase information in ultrasonic backscatter for in vivo liver analysisUltrasound in Medicine & Biology, 1998
- Statistical characterization of diffuse scattering in ultrasound imagesIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- An adaptive weighted median filter for speckle suppression in medical ultrasonic imagesIEEE Transactions on Circuits and Systems, 1989
- Some fundamental properties of speckle*Journal of the Optical Society of America, 1976
- A model for non-Rayleigh sea echoIEEE Transactions on Antennas and Propagation, 1976