Data-acquisition system for measurement of thermal diffusivity and propagation properties of thermal waves by a non-steady-state method
Authors:
- 1 August 1986
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 57 (8) , 1627-1632
Abstract
We describe an implementation of the experimental technique used for measuring thermal diffusivity and propagation properties of thermal waves by means of a non-steady-state method. The values of the phase shift and amplitude ratio of the temperature oscillation at two fixed points of the specimen are obtained directly by a data-acquisition and processing system which uses a properly interfaced microcomputer. Moreover, in order to reduce substantially the constant temperature gradient along the specimen, the heat oscillating source was replaced by a Peltier device supplied with an alternating current.Keywords
This publication has 3 references indexed in Scilit:
- Thermal Diffusivity of Metals at High TemperaturesJournal of Applied Physics, 1954
- XXIII. On the conducting-power of copper and iron for heat at different temperaturesJournal of Computers in Education, 1863
- XVII. New method of determining the thermal conductibility of bodiesJournal of Computers in Education, 1863