Effect of printing orientation on mechanical properties of 3D-printed orthodontic aligners
Open Access
- 26 January 2024
- journal article
- research article
- Published by Springer Nature in Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie
- Vol. 86 (4) , 226-233
Abstract
Purpose: The purpose of this study was to assess differences in the fundamental mechanical properties of resin-made three-dimensional (3D) printed orthodontic aligners according to the printing orientation. Methods: Twenty resin 3D-printed dumbbell-shaped specimens and 20 orthodontic aligners were fabricated and postcured in nitrogen. Half of the specimens and aligners were built in horizontal (H), the other half in vertical (V) directions. The dumbbell-shaped specimens were loaded in a tensile testing machine, while parts of the aligners were embedded in acrylic resin, ground, polished, and then underwent instrumented indentation testing (IIT). Mechanical properties that were assessed included the yield strength (YS), breaking strength (BS), plastic strain (ε), Martens hardness (HM), indentation modulus (EIT), elastic index (ηIT), and indentation relaxation (RIT). Data were analyzed statistically with independent t‑tests or Mann–Whitney tests at α = 5%. Results: No significant differences were found between specimens or aligners printed either in a horizontal or a vertical direction (P > 0.05 in all instances). Overall, the 3D-printed aligners showed acceptable mechanical propertied in terms of YS (mean 19.2 MPa; standard deviation [SD] 1.7 MPa), BS (mean 19.6 MPa; SD 1.2 MPa), ε (mean 77%; SD 11%), HM (median 89.0 N/mm2; interquartile range [IQR] 84.5–90.0 NN/m2), EIT (median 2670.5 MPa; IQR 2645.0–2726.0 MPa), ηIT (median 27.5%; IQR 25.9–28.1%), and RIT (mean 65.1%; SD 3.5%). Conclusion: Printing direction seemed to have no effect on the mechanical properties of 3D-printed resin aligners, which are promising for orthodontic use.Keywords
Funding Information
- University of Zurich
This publication has 33 references indexed in Scilit:
- Stress relaxation properties of four orthodontic aligner materials: A 24-hour in vitro studyThe Angle Orthodontist, 2016
- Scanning electron microscopy analysis of the growth of dental plaque on the surfaces of removable orthodontic aligners after the use of different cleaning methodsClinical, Cosmetic and Investigational Dentistry, 2015
- Do the mechanical and chemical properties of InvisalignTMappliances change after use? A retrieval analysisEuropean Journal of Orthodontics, 2015
- Chemical and mechanical characteristics of contemporary thermoplastic orthodontic materialsAustralasian Orthodontic Journal, 2015
- Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliancesThe Angle Orthodontist, 2013
- Dynamic stress relaxation of orthodontic thermoplastic materials in a simulated oral environmentDental Materials Journal, 2013
- Effects of environmental conditions, aging, and build orientations on the mechanical properties of ASTM type I specimens manufactured via stereolithographyRapid Prototyping Journal, 2012
- Editor's Summary and Q&AAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2009
- Effects of build orientation on tensile strength for stereolithography-manufactured ASTM D-638 type I specimensThe International Journal of Advanced Manufacturing Technology, 2009
- Review of instrumented indentationJournal of Research of the National Institute of Standards and Technology, 2003