The Value of DDD Pacing in Patients with an Implantable Cardioverter Defibrillator
Authors:
- 1 January 1997
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 20 (1) , 177-181
Abstract
Although the beneficial effects of DDD pacing are well known, currently available ICDs provide only fixed rate ventricular antibradycardia pacing. In a consecutive series of 139 patients with ICDs, we have analyzed the need for antibradycardia pacing and the indications for DDD pacing. We also report our initial experience with the Defender 9001 (ELA Medical, France) DDD‐ICD. Out of 139 patients, 25 (18%) were in need of antibradycardia pacing. Ten patients already had a pacemaker at the time of ICD implantation and ten other patients had a conventional pacemaker indication at that time. Five patients became pacemaker dependent during a follow‐up of 20 ± 8 months. The disorders necessitating pacemaker therapy were high degree AV conduction disturbances in 72%, sick sinus syndrome in 12%, and AF with a slow ventricular response in 16% of patients. Based upon current indications, DDD pacing was indicated in 20 (80%) of 25 patients. The Defender 9001 DDD‐ICD (ELA Medical) was used in two patients with ischemic cardiomyopathy and pacemaker syndrome with VVI pacing. Cardiac output during DDD pacing increased by 36% in one patient with an increase in VO2 max during exercise of 29%. The other patient showed an increase in cardiac output of 50% with DDD pacing, and, while unable to exercise with VVI pacing, had a VO2max of 24 mL/kg per minute during DDD pacing. Up to 18% of our ICD patients are in need of antibradycardia pacing. Of these pacemaker dependent patients, 80% have an indication for DDD pacing. Our first clinical experience with a DDD‐ICD confirms the hemodynamic benefit of AV synchronous pacing in ICD patients with pacemaker syndrome.Keywords
This publication has 8 references indexed in Scilit:
- Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakersAmerican Heart Journal, 1995
- Deleterious effects of long-term single-chamber ventricular pacing in patients with sick sinus syndrome: The hidden benefits of dual-chamber pacingJACC, 1992
- Guidelines for implantation of cardiac pacemakers and antiarrhythmia devicesJACC, 1991
- Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillatorsJACC, 1990
- Long-term pacing in sinus node disease: Effects of stimulation mode on cardiovascular morbidity and mortalityAmerican Heart Journal, 1988
- Developments for Future Implantable Cardioverters and DefibrillatorsPacing and Clinical Electrophysiology, 1987
- The Natural History of Sick Sinus SyndromePacing and Clinical Electrophysiology, 1986
- Hemodynamic effects of acute atrioventricular sequential pacing in patients with left ventricular dysfunctionThe American Journal of Cardiology, 1982