Control of plastic extension in fatigue tests

Authors:

Abstract
An electronic device is described which is used to adapt a servo-controlled mechanical testing machine for fatigue tests at constant plastic strain amplitude. Implementation involves no alterations to the existing servo loop and is therefore quite simple. Also described is a modification to the basic device which permits precise strain control at greatly increased speeds and also reduces sensitivity to phase shifts (e.g. those caused by changes in the servo loop gain resulting from hardening of the specimen). In experimental use with a general purpose testing machine (Instron 1253) the device was able to maintain an oscillating plastic extension of +or-30 mu m to within 1 mu m at frequencies up to 30 Hz.