Spatial Accuracy of a Clinically Established Noninvasive Electrocardiographic Imaging System for the Detection of Focal Activation in an Intact Porcine Model
Authors:
- 1 November 2019
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 12 (11) , e007570
Abstract
Background: Noninvasive electrocardiographic imaging (ECGi) is used clinically to map arrhythmias before ablation. Despite its clinical use, validation data regarding the accuracy of the system for the identification of arrhythmia foci is limited. Methods: Nine pigs underwent closed-chest placement of endocardial fiducial markers, computed tomography, and pacing in all cardiac chambers with ECGi acquisition. Pacing location was reconstructed from biplane fluoroscopy and registered to the computed tomography using the fiducials. A blinded investigator predicted the pacing location from the ECGi data, and the distance to the true pacing catheter tip location was calculated. Results: A total of 109 endocardial and 9 epicardial locations were paced in 9 pigs. ECGi predicted the correct chamber of origin in 85% of atrial and 92% of ventricular sites. Lateral locations were predicted in the correct chamber more often than septal locations (97% versus 79%, P =0.01). Absolute distances in space between the true and predicted pacing locations were 20.7 (13.8–25.6) mm (median and [first–third] quartile). Distances were not significantly different across cardiac chambers. Conclusions: The ECGi system is able to correctly identify the chamber of origin for focal activation in the vast majority of cases. Determination of the true site of origin is possible with sufficient accuracy with consideration of these error estimates.Keywords
This publication has 30 references indexed in Scilit:
- Validation of Novel 3-Dimensional Electrocardiographic Mapping of Atrial Tachycardias by Invasive Mapping and AblationJACC, 2013
- Inverse Solution Mapping of Epicardial PotentialsCirculation: Arrhythmia and Electrophysiology, 2012
- Noninvasive Characterization of Epicardial Activation in Humans With Diverse Atrial Fibrillation PatternsCirculation, 2010
- The three‐dimensional arrangement of the myocytes in the ventricular wallsClinical Anatomy, 2008
- Three-Dimensional Ultrasound for Image-Guided Mapping and InterventionCirculation: Arrhythmia and Electrophysiology, 2008
- Epicardial and intramural excitation during ventricular pacing: effect of myocardial structureAmerican Journal of Physiology-Heart and Circulatory Physiology, 2008
- Noninvasive Electrocardiographic Imaging (ECGI): Comparison to intraoperative mapping in patientsHeart Rhythm, 2005
- Effects of experimental and modeling errors on electrocardiographic inverse formulationsIEEE Transactions on Biomedical Engineering, 2003
- Virtual cathode effects during stimulation of cardiac muscle. Two-dimensional in vivo experiments.Circulation Research, 1991
- Direct Current Make and Break Thresholds for Pacemaker Electrodes on the Canine VentricleCirculation Research, 1970