Influence of Sex, BMI, and Skin Color on the Accuracy of Non-Invasive Cuffless Photoplethysmography-Based Blood Pressure Measurements
Authors:
Open Access
- 29 June 2022
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Physiology
- Vol. 13, 911544
Abstract
Vital signs obtained by photoplethysmography-based devices might be influenced by subcutaneous fat and skin color. This observational comparison study aimed to test the accuracy of blood pressure (BP) measurements between a photoplethysmography-based device and cuff-based BP device in ambulatory individuals, coming for a routine BP checkup. Systolic BP (SBP) and diastolic BP (DBP) measurements were stratified based on sex, BMI (2), and skin color (types 1–3 and 4–6 by the Fitzpatrick scale). A total of 1548 measurements were analyzed. Correlations of SBP and DBP between the devices among males/females were between 0.914–0.987 (p < 0.001), and Bland-Altman analysis showed a bias of less than 0.5 mmHg for both sexes. Correlations of SBP and DBP between the devices among BMI groups were between 0.931–0.991 (p < 0.001), and Bland-Altman analysis showed a bias of less than 1 mmHg for all. Correlations of SBP and DBP between the devices among the skin color groups were between 0.936–0.983 (p < 0.001), and Bland-Altman analysis showed a bias of less than 1 mmHg for all. This study shows similar and high agreements between BP measurements obtained using a PPG-based non-invasive cuffless BP device and a cuff-based BP device across sex, BMI, and skin color groups.Keywords
This publication has 26 references indexed in Scilit:
- Why are sex and gender important to basic physiology and translational and individualized medicine?American Journal of Physiology-Heart and Circulatory Physiology, 2014
- Blood pressure measurement in very obese patients: a challenging problemJournal Of Hypertension, 2011
- Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: The Effects of Oximeter Probe Type and GenderAnesthesia & Analgesia, 2007
- Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low SaturationAnesthesiology, 2005
- Effect of Skin Pigmentation on Pulse Oximetry Accuracy in the Emergency DepartmentAcademic Emergency Medicine, 1998
- Reliability of Noninvasive Oximetry in Black Subjects during Exercise and HypoxiaAmerican Review of Respiratory Disease, 1991
- Reliability of Pulse Oximetry in Titrating Supplemental Oxygen Therapy in Ventilator-Dependent PatientsChest, 1990
- Skin Color and Ear OximetryChest, 1989
- The validity and practicality of sun-reactive skin types I through VIArchives of Dermatology, 1988
- The Validity and Practicality of Sun-Reactive Skin Types I Through VIArchives of Dermatology, 1988