Abstract
During aeromedical patient transport, acoustic noise levels in excess of 90 dB make auscultation with a stethoscope virtually impossible. Existing approaches to address this noise in literature have relied on the use of either passive shielding of the stethoscope sensing element or active noise cancellation (ANC) with the use of additional reference microphone sensors. These solutions have achieved inadequate performance. In this paper, custom instrumented stethoscopes are used to obtain structural vibration and acoustic data from a UH-60 Black Hawk helicopter. The data show that the structural vibrations contribute significantly more to stethoscope signals than air-borne acoustic noise. Therefore, a novel adaptive acoustic control system using a reference accelerometer is pursued. By varying the levels of acoustic and vibrational noise sources in a controlled laboratory environment, the benefits and limitations of microphone or accelerometer-based active noise control are investigated and compared. Experimentally simulating the field acoustic and vibration characteristics of the UH-60 Black Hawk helicopter, the accelerometer-based ANC system is shown to provide significantly superior performance. The accelerometer-based ANC system yields an average 25-dB reduction in ambient noise levels over the entire relevant frequency range, thereby enabling successful auscultation for the first time in this military aircraft environment.Keywords
Funding Information
- Department of the Army and 3M's Infection Prevention Division
- U.S. Army Medical Research Acquisition; Cooperative Agreement (W81XWH-10-1-1051)
This publication has 32 references indexed in Scilit:
- In flight auscultation: comparison of electronic and conventional stethoscopesThe American Journal of Emergency Medicine, 2011
- The Effect of Noise in the Emergency DepartmentAcademic Emergency Medicine, 2005
- Assessment of Breath Sounds During Ambulance TransportAnnals of Emergency Medicine, 1997
- Noise-reducing stethoscopeThe Journal of the Acoustical Society of America, 1996
- Acoustic properties of the normal chestEuropean Respiratory Journal, 1995
- Normalized data nonlinearities for LMS adaptationIEEE Transactions on Signal Processing, 1994
- In-flight experiments on the active control of propeller-induced cabin noiseJournal of Sound and Vibration, 1990
- The Hospital-Based Helicopter: A Threat to Hearing?JAMA Otolaryngology–Head & Neck Surgery, 1985
- Hearing in an ambulance paramedic populationAnnals of Emergency Medicine, 1980
- The effect of background noise on cardiac auscultationAmerican Heart Journal, 1956