Electrocardiogram Provides a Continuous Heart Rate Faster Than Oximetry During Neonatal Resuscitation
Open Access
- 1 November 2012
- journal article
- research article
- Published by American Academy of Pediatrics (AAP) in Pediatrics
- Vol. 130 (5) , e1177-e1181
Abstract
OBJECTIVE: To compare the time required to obtain a continuous audible heart rate signal from an electrocardiogram (ECG) monitor and pulse oximeter (PO) in infants requiring resuscitation. METHODS: Infants who had both ECG and PO placed during resuscitation were analyzed using video and analog recordings. The median times from arrival until the ECG electrodes and PO sensor were placed, and the time to achieve audible tones from the devices, were compared. RESULTS: Forty-six infants had ECG and PO data. Thirty infants were very low birth weight (23–30 weeks). There was a difference in the median total time to place either device (26 vs 38 seconds; P = .04), and a difference (P < .001) in the time to achieve an audible heart rate signal after ECG lead (2 seconds) versus PO probe (24 seconds) placement. In infants weighing >1500 g (n = 16), the median time (interquartile range) to place the ECG was 20 seconds (14–43) whereas the time to place the PO was 36 seconds (28–56) (P = .74). The median times (interquartile range) to acquire a signal from the ECG and PO were 4 seconds (1–6) and 32 seconds (15–40, P = .001), respectively. During the first minutes of resuscitation, 93% of infants had an ECG heart rate compared with only 56% for PO. CONCLUSIONS: Early application of ECG electrodes during infant resuscitation can provide the resuscitation team with a continuous audible heart rate, and its use may improve the timeliness of appropriate critical interventions.Keywords
This publication has 7 references indexed in Scilit:
- Electrocardiogram shows reliable heart rates much earlier than pulse oximetry during neonatal resuscitationPediatrics International, 2011
- Delivery Room Intervention: Improving the OutcomeClinics in Perinatology, 2010
- Accuracy of Pulse Oximetry Measurement of Heart Rate of Newborn Infants in the Delivery RoomThe Journal of Pediatrics, 2008
- Accuracy of pulse oximetry in assessing heart rate of infants in the neonatal intensive care unitJournal of Paediatrics and Child Health, 2007
- Accuracy of clinical assessment of infant heart rate in the delivery roomResuscitation, 2006
- Premedication for Nonemergent Neonatal Intubations: A Randomized, Controlled Trial Comparing Atropine and Fentanyl to Atropine, Fentanyl, and MivacuriumPediatrics, 2006
- Feasibility of and Delay in Obtaining Pulse Oximetry during Neonatal ResuscitationThe Journal of Pediatrics, 2005