21 Tesla Fourier Transform Ion Cyclotron Resonance Mass Spectrometer: A National Resource for Ultrahigh Resolution Mass Analysis
- 20 June 2015
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 26 (9) , 1626-1632
Abstract
We describe the design and initial performance of the first 21 tesla Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer. The 21 tesla magnet is the highest field superconducting magnet ever used for FT-ICR and features high spatial homogeneity, high temporal stability, and negligible liquid helium consumption. The instrument includes a commercial dual linear quadrupole trap front end that features high sensitivity, precise control of trapped ion number, and collisional and electron transfer dissociation. A third linear quadrupole trap offers high ion capacity and ejection efficiency, and rf quadrupole ion injection optics deliver ions to a novel dynamically harmonized ICR cell. Mass resolving power of 150,000 (m/Δm 50% ) is achieved for bovine serum albumin (66 kDa) for a 0.38 s detection period, and greater than 2,000,000 resolving power is achieved for a 12 s detection period. Externally calibrated broadband mass measurement accuracy is typically less than 150 ppb rms, with resolving power greater than 300,000 at m/z 400 for a 0.76 s detection period. Combined analysis of electron transfer and collisional dissociation spectra results in 68% sequence coverage for carbonic anhydrase. The instrument is part of the NSF High-Field FT-ICR User Facility and is available free of charge to qualified users. Graphical Abstract ᅟ ᅟThis publication has 34 references indexed in Scilit:
- Performance evaluation of a dual linear ion trap-Fourier transform ion cyclotron resonance mass spectrometer for proteomics researchJournal of Proteomics, 2013
- High Resolution Mass SpectrometryAnalytical Chemistry, 2012
- Initial Experimental Characterization of a New Ultra-High Resolution FTICR Cell with Dynamic HarmonizationJournal of the American Society for Mass Spectrometry, 2011
- Fourier transform ion cyclotron resonance cell with dynamic harmonization of the electric field in the whole volume by shaping of the excitation and detection electrode assemblyRapid Communications in Mass Spectrometry, 2010
- Dual-Pressure Linear Ion Trap Mass Spectrometer Improving the Analysis of Complex Protein MixturesAnalytical Chemistry, 2009
- High-Performance Mass Spectrometry: Fourier Transform Ion Cyclotron Resonance at 14.5 TeslaAnalytical Chemistry, 2008
- A two-dimensional quadrupole ion trap mass spectrometerJournal of the American Society for Mass Spectrometry, 2002
- Fourier transform ion cyclotron resonance mass spectrometry: A primerMass Spectrometry Reviews, 1998
- Advantages of High Magnetic Field for Fourier Transform Ion Cyclotron Resonance Mass SpectrometryRapid Communications in Mass Spectrometry, 1996
- Micro-electrospray mass spectrometry: Ultra-high-sensitivity analysis of peptides and proteinsJournal of the American Society for Mass Spectrometry, 1994