Abstract
Background Fractures of the distal radius represent the most common fractures in adults. Volar locked plating has become a popular method for treating these fractures, but has been subject to several shortcomings in osteoporotic bone, such as loss of reduction and screw purchase. In order to overcome these shortcomings, cement augmentation has been proposed. Methods AO-type 23-A3.3 fractures were made in 8 pairs of fresh frozen osteoporotic cadaveric radial bones. All specimens were treated with volar plating, and divided into cement augmentation or non-augmentation groups (n = 8/group). Constructs were tested dynamically and load to failure, construct-stiffness, fracture gap movement and screw cutting distance were measured. Results Cement augmentation resulted in a significant increase in cycles and load to failure, as well as construct stiffness at loads higher than 325 N. When compared to the non-augmented group, fracture gap movement decreased significantly at this load and higher, as did screw cutting distance at the holes of the ulnar column. The cycles to failure depend on the BMD in the distal region of the radius. Conclusion Cement augmentation improves biomechanical properties in volar plating of the distal radius.Keywords
This publication has 23 references indexed in Scilit:
- Biomechanical effect of bone cement augmentation on rotational stability and pull-out strength of the Proximal Femur Nail Antirotation™Injury, 2011
- Unstable Distal Radius Fractures in the Elderly Patient—Volar Fixed-Angle Plate Osteosynthesis Prevents Secondary Loss of ReductionJournal of Trauma: Injury, Infection & Critical Care, 2010
- Locked volar plating for unstable distal radial fractures: Clinical and radiological outcomesInjury, 2010
- A novel non-bridging external fixator construct versus volar angular stable plating for the fixation of intra-articular fractures of the distal radius—A biomechanical studyInjury, 2010
- Does a Volar Locking Plate Provide Equivalent Stability as a Dorsal Nonlocking Plate in a Dorsally Comminuted Distal Radius Fracture?Journal of Orthopaedic Trauma, 2008
- Grenzen der palmaren winkelstabilen Plattenosteosynthese bei instabilen distalen RadiusfrakturenHandchirurgie · Mikrochirurgie · Plastische Chirurgie, 2007
- Functional Outcome and Complications After Volar Plating for Dorsally Displaced, Unstable Fractures of the Distal RadiusThe Journal of Hand Surgery, 2006
- Die palmare winkelstabile Plattenosteosynthese bei distalen RadiusextensionsfrakturenAktuelle Traumatologie, 2001
- Fractures of the Distal Radius in a Norwegian CityScandinavian Journal of Plastic and Reconstructive Surgery and Hand Surgery, 1995
- TREATMENT AND PROGNOSIS OF COLLES' FRACTUREThe Lancet, 1963