Accuracy and precision of blood pressure determination with the Finapres: an overview using re-sampling statistics
Authors:
Open Access
- 1 June 1998
- journal article
- research article
- Published by Springer Nature in Journal of Human Hypertension
- Vol. 12 (6) , 403-409
Abstract
The Finapres non-invasive blood pressure (BP) monitor uses the method of Penaz to indirectly record the arterial waveform; studies on its accuracy have suggested little systematic bias vs intra-arterial pressure (IAP) but substantial variability. Inconsistency between studies, in respect of the magnitude, direction and variance of bias, was described in the validation studies against the direct IAP. We have employed a novel re-sampling statistical method to combine the data from 20 published studies; a robust overall estimate of the accuracy and precision of the Finapres was thereby obtained. Based on 449 patients and 4490 re-samples, the average Finapres systolic bias (IAP - Finapres) was 2.2 mm Hg (s.d. ±12.4) with limits of agreement (bias ±2 s.d.) of −22.6 and 26.9 mm Hg. The average precision was 12.1 mm Hg (s.d. ±8.4). The Finapres diastolic bias was −0.3 mm Hg (s.d. ±7.9) with the limits of agreement of −16.1 and 15.5 mm Hg. The average precision was 7.6 mm Hg (s.d. ±5.3). The average Finapres mean arterial pressure bias was 2.1 mm Hg (s.d. ±8.6) with precision of 7.6 mm Hg (s.d. ±5.3). The calculated percentage of Finapres systolic values expected to fall within ±5 or ±10 of the direct intra-arterial pressure was 35.9% and 73.1%, respectively. The calculated precision of the Finapres systolic pressure between 0–5 mm Hg was 1.6% and between 0–10 mm Hg 36.4%. The comparable values for Finapres diastolic BP for accuracy were 63.5% and 92.8% and for precision 23.1% and 79.2%. The Finapres device can provide an accurate estimate of diastolic and mean arterial pressure compared with the intra-arterial record; the apparent inaccuracy of the Finapres systolic pressure may have a physiological explanation. When the Finapres device is used in experimental or in clinical situations, then calibration against a reliable reference arterial pressure is desirable to obviate the possibility of an ‘offset’ error.Keywords
This publication has 24 references indexed in Scilit:
- Reliability of blood pressure determination with the Finapres with altered physiological states or pharmacodynamic conditionsClinical Autonomic Research, 1997
- Evaluation of non-invasive blood pressure measurement by the Finapres method at rest and during dynamic exercise in subjects with cardiovascular insufficiencyClinical Autonomic Research, 1994
- Comparative accuracies of a finger blood pressure monitor and an oscillometric blood pressure monitorJournal of Clinical Monitoring and Computing, 1991
- Continuous finger arterial pressure: Utility in the cardiovascular laboratoryClinical Autonomic Research, 1991
- Evaluation of a continuous noninvasive blood pressure monitor in obstetric patients undergoing spinal anesthesiaJournal of Clinical Monitoring and Computing, 1989
- Noninvasive continuous blood pressure measurement from the finger: Optimal measurement conditions and factors affecting reliabilityJournal of Clinical Monitoring and Computing, 1987
- Effects of peripheral vasoconstriction on the measurement of blood pressure in a fingerCardiovascular Research, 1985
- Evaluation of two prototype devices producing noninvasive, pulsatile, calibrated blood pressure measurement from a fingerJournal of Clinical Monitoring and Computing, 1985
- Evaluation of the Penàz servo-plethysmo-manometer for the continuous, non-invasive measurement of finger blood pressureBasic Research in Cardiology, 1984
- The arterial pulse in health and diseaseAmerican Heart Journal, 1971