Effects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans
Open Access
- 8 September 2015
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 16 (5) , 87-105
Abstract
This study was aimed at investigating delivery quality assurance (DQA) discrepancies observed for helical tomotherapy plans. A selection of tomotherapy plans that initially failed the DQA process was chosen for this investigation. These plans failed the fluence analysis as assessed using gamma criteria (3%, 3 mm) with radiographic film. Each of these plans was modified (keeping the planning constraints the same), beamlets rebatched and reoptimized. By increasing and decreasing the modulation factor, the fluence in a circumferential plane as measured with a diode array was assessed. A subset of these plans was investigated using varied pitch values. Metrics for each plan that were examined were point doses, fluences, leaf opening times, planned leaf sinograms, and uniformity indices. In order to ensure that the treatment constraints remained the same, the dose‐volume histograms (DVHs) of all the modulated plans were compared to the original plan. It was observed that a large increase in the modulation factor did not significantly improve DVH uniformity, but reduced the gamma analysis pass rate. This also increased the treatment delivery time by slowing down the gantry rotation speed which then increases the maximum to mean non‐zero leaf open time ratio. Increasing and decreasing the pitch value did not substantially change treatment time, but the delivery accuracy was adversely affected. This may be due to many other factors, such as the complexity of the treatment plan and site. Patient sites included in this study were head and neck, right breast, prostate, abdomen, adrenal, and brain. The impact of leaf timing inaccuracies on plans was greater with higher modulation factors. Point‐dose measurements were seen to be less susceptible to changes in pitch and modulation factors. The initial modulation factor used by the optimizer, such that the TPS generated ‘actual’ modulation factor within the range of 1.4 to 2.5, resulted in an improved deliverable plan.PACS number: 87.55.‐x, 87.55.Qr, 87.55.D‐Keywords
This publication has 15 references indexed in Scilit:
- Evaluation of dose coverage to target volume and normal tissue sparing in the adjuvant radiotherapy of gastric cancers: 3D-CRT compared with dynamic IMRTBiomedical Imaging and Intervention Journal, 2010
- Use of Normal Tissue Complication Probability Models in the ClinicInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Treatment Planning to Improve Delivery Accuracy and Patient Throughput in Helical TomotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Helical Tomotherapy Quality AssuranceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- A treatment planning study comparing whole breast radiation therapy against conformal, IMRT and tomotherapy for accelerated partial breast irradiationRadiotherapy and Oncology, 2007
- The helical tomotherapy thread effectMedical Physics, 2005
- Quality assurance of a helical tomotherapy machinePhysics in Medicine & Biology, 2004
- Reduction in Radiation Dose to Lung and Other Normal Tissues Using Helical Tomotherapy to Treat Lung Cancer, in Comparison to Conventional Field ArrangementsAmerican Journal of Clinical Oncology, 2003
- High-dose intensity modulated radiation therapy for prostate cancer: early toxicity and biochemical outcome in 772 patientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- Megavoltage CT on a tomotherapy systemPhysics in Medicine & Biology, 1999