Dose Reduction in Helical CT: Dynamically Adjustable z-Axis X-Ray Beam Collimation
Authors:
- 1 January 2010
- journal article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 194 (1) , W49-W55
Abstract
OBJECTIVE. The purpose of this study was to measure the dose reduction achieved with dynamically adjustable z-axis collimation.MATERIALS AND METHODS. A commercial CT system was used to acquire CT scans with and without dynamic z-axis collimation. Dose reduction was measured as a function of pitch, scan length, and position for total incident radiation in air at isocenter, accumulated dose to the center of the scan volume, and accumulated dose to a point at varying distances from a scan volume of fixed length. Image noise was measured at the beginning and center of the scan.RESULTS. The reduction in total incident radiation in air at isocenter varied between 27% and 3% (pitch, 0.5) and 46% and 8% (pitch, 1.5) for scan lengths of 20 and 500 mm, respectively. Reductions in accumulated dose to the center of the scan were 15% and 29% for pitches of 0.5 and 1.5 for 20-mm scans. For scan lengths greater than 300 mm, dose savings were less than 3% for all pitches. Dose reductions 80 mm or farther from a 100-mm sc...Keywords
This publication has 9 references indexed in Scilit:
- Minimization of over-ranging in helical volumetric CT via hybrid cone beam image reconstruction-Benefits in dose efficiencyMedical Physics, 2008
- Influence of overscanning on normalized effective doses calculated for pediatric patients undergoing multidetector CT examinationsMedical Physics, 2007
- Energy imparted‐based estimates of the effect of overscanning on adult and pediatric patient effective doses from multi‐slice computed tomographyMedical Physics, 2007
- Overranging in Multisection CT: Quantification and Relative Contribution to Dose—Comparison of Four 16-Section CT ScannersRadiology, 2007
- Female Breast Radiation Exposure During CT Pulmonary AngiographyAmerican Journal of Roentgenology, 2005
- The effect of overscanning on patient effective dose from multidetector helical computed tomography examinationsMedical Physics, 2005
- Primary radiation outside the imaged volume of a multislice helical CT scanBritish Journal of Radiology, 2002
- Monte Carlo assessment of computed tomography dose to tissue adjacent to the scanned volumeMedical Physics, 2000
- Computed tomography scanning with simultaneous patient translationMedical Physics, 1990