Abstract
In this paper, an optimal design for Nuclear Magnetic Resonance (NMR) and Magnetic Resonance Imaging (MRI)-based multi-phase flow meter (MPFM) is presented. The overall apparatus consists of an electromagnetic coil surrounded by 12 rings of Halbach arrays of 12 magnets each. Both the coil and the magnets arrays were simultaneously and optimally designed using a 3D Finite Element (FEM) package and Particle Swarm Optimization (PSO) algorithm. The aim of the design is to uniformly prepolarize the hydrogen spins of the atoms composing the flowing fluid, with a predefined static magnetic field, before they enter into the measurement area within which they are homogenously polarized with a perpendicular AC magnetic field. Cheap and light weight hardware are the other two design parameters of the design. Results of extensive 3D simulations of the design indicate that an optimized and homogenous static magnetic field distribution could be achieved within an area of 40 mm diameter and 606 mm length. In addition, using a multi-turn coil of 6 cm diameter and 2000 turns an AC magnetic field of 13 mT amplitude and 112 ppm homogeneity could be achieved; which is enough to handle fluids flowing at speed of up to 2m/s. Experimental validation which was done using newly constructed two Halbach arrays of cuboid and trapezoid magnet elements respectively indicate a good match with FEM simulations.
Funding Information
  • Abu Dhabi National Oil Company

This publication has 12 references indexed in Scilit: