Combined use of Finite Element and Equivalent Circuit Modeling for System-Level Simulation of Integrated Capacitive Micromachined Ultrasonic Transducers (CMUT)
Authors:
- 7 September 2020
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
This paper proposes the combined use of Finite Element Modeling (FEM) and Equivalent Circuit Modeling (ECM) to describe the nonlinear electromechanical and acoustic behavior of a CMUT in an integrated circuit simulation environment. CMUT ECM lumped parameters are computed by FEM simulations. The proposed method is implemented in LTSpice, and validated by comparing ECM and FEM static, small-signal harmonic, and large-signal transient simulation results. The model was then applied to the performance assessment of the CMUT coupled to a realistic integrated transceiver configuration by computing TX efficiency and noise figure. The proposed approach allows accurate large-signal analysis of CMUTs interfaced to ultrasound front-end circuits by considering non-ideal characteristics of the system, such as the pulser nonlinearities.Keywords
This publication has 6 references indexed in Scilit:
- A Comparative Analysis of CMUT Receiving Architectures for the Design Optimization of Integrated Transceiver Front EndsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2017
- Acoustic reflectivity minimization in Capacitive Micromachined Ultrasonic Transducers (CMUTs)Ultrasonics, 2017
- A volumetric CMUT-based ultrasound imaging system simulator with integrated reception and μ-beamforming electronics modelsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2014
- A CMUT probe for medical ultrasonography: from microfabrication to system integrationIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2012
- Micromachined Ultrasonic Transducers2008
- Enhanced echographic images obtained improving the membrane structural layer of the cMUT probe2005