Comparison of Conventional Technique with Suture Fixation and Subperiosteal Tight Pocket Technique on Revision Cochlear Implantation Rate
Open Access
- 29 July 2024
- journal article
- research article
- Published by Caleo in The Journal of International Advanced Otology
- Vol. 20 (4) , 301-305
Abstract
Background: Several fixation methods have been described to secure the cochlear implant’s receiver/stimulator, but the optimal stabilization technique is still being debated. The aim of this study was to compare the conventional technique with suture fixation to the subperiosteal tight pocket technique in terms of revision cochlear implantation rate. Methods: A retrospective review was conducted on the medical records of 649 patients who underwent cochlear implantation. The study participants were divided into different groups regarding the applied surgery technique. The relationship between the fixation technique, revision rates, and the cause of revisions related to techniques was investigated. Results: The overall revision rate was 2.9% (19 out of 649). There were 14 (3.5%) and 5 (2%) revision implantations in the subperiosteal tight pocket and conventional technique groups, respectively. The incidence of device failure was 2.5%, and it constituted the primary cause for revision surgery in both groups. Even though patients who had the subperiosteal tight pocket technique had a much higher rate of device failure, the results indicate that there was no significant difference between the groups, as evidenced by a P-value of .12. Conclusion: The conventional and subperiosteal tight pocket techniques can both be safely preferred with low revision rates in patients undergoing cochlear implantation. Cite this article as: Ceylan ME, Zorlu ME, Çorakçı O, Akı ES, Aksoy Yıldırım G, Dalgic A. Comparison of conventional technique with suture fixation and subperiosteal tight pocket technique on revision cochlear implantation rate. J Int Adv Otol. 2024;20(4):301-305.Keywords
This publication has 19 references indexed in Scilit:
- The effect of cochlear implant bed preparation and fixation technique on the revision cochlear implantation rateThe Journal of Laryngology & Otology, 2018
- Evaluation of receiver–stimulator migration in cochlear implantation using the subperiosteal pocket technique: a prospective clinical studyThe Journal of Laryngology & Otology, 2017
- 228 Cases of Cochlear Implant Receiver‐Stimulator Placement in a Tight Subperiosteal Pocket without FixationOtolaryngology–Head and Neck Surgery, 2015
- Spontaneous Bone Bed Formation in Cochlear Implantation Using the Subperiosteal Pocket TechniqueOtology & Neurotology, 2014
- Cochlear implant fixation using a subperiosteal tight pocket without either suture or bone‐recess techniqueThe Laryngoscope, 2014
- Cochlear Implant Failure, Revision, and ReimplantationOtology & Neurotology, 2013
- Minimal Access Cochlear Implant FixationOtology & Neurotology, 2012
- Flap Outcome Using “C” Shaped and “New” Incisions in Pediatric Cochlear Implantation2012
- Fixation of cochlear implants: an evidence-based review of literature.2012
- Cochlear Implant Soft Failures Consensus Development Conference StatementOtology & Neurotology, 2005