Biomechanical Response under Stress-Controlled Tension-Tension Fatigue of a Novel Carbon Fiber/Epoxy Intramedullary Nail for Femur Fractures
- 11 May 2020
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 80 (1) , 26-32
Abstract
Metallic intramedullary nails are the "gold standard" implant for repairing femur shaft fractures. However, their rigidity may eliminate axial micromotion at the fracture (causing delayed healing) and they may carry too much load relative to the femur (causing "stress shielding"). Consequently, some researchers have proposed fiber-reinforced composite nails, but only one evaluated cyclic fatigue performance. Therefore, this study assessed the cyclic fatigue response of a carbon fiber/epoxy nail with a novel ply stacking sequence of [02/-45/452/-45/0/-45/452/-452/452/-45/902] previously developed by the present authors. Nails were cyclically loaded in tension-tension at 5 Hz with a stress ratio of R=0.1 from 30% - 85% of the material's ultimate tensile strength (UTS). Thermographic stress analysis, rather than conventional fatigue testing, was used to obtain high cycle fatigue strength (HCFS), below which the nail can be cyclically loaded indefinitely without damage. Also, the mechanical test machine's built-in load cell and an extensometer were used to create stress-strain curves, from which the change in static EO and dynamic E* moduli were obtained. Results showed that HCFS was 70.3% of UTS (or about 283 MPa), while EO and E* remained at 42 GPa without any dRegradation during testing. The current nail shows potential for clinical use.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada
This publication has 43 references indexed in Scilit:
- Carbon fiber reinforced PEEK Optima—A composite material biomechanical properties and wear/debris characteristics of CF-PEEK composites for orthopedic trauma implantsJournal of the Mechanical Behavior of Biomedical Materials, 2013
- Determination of the fatigue limit of a carbon/epoxy composite using thermographic analysisStructural Control and Health Monitoring, 2011
- Effect of mechanical stability on fracture healing — an updateInjury, 2007
- Assessment of Carbon Fibre Composite Fracture Fixation Plate Using Finite Element AnalysisAnnals of Biomedical Engineering, 2006
- The course of bone healing is influenced by the initial shear fixation stabilityJournal of Orthopaedic Research, 2005
- Stiffness and strength design of composite bone platesComposites Science and Technology, 2005
- Feasibility of knitted carbon/PEEK composites for orthopedic bone platesBiomaterials, 2004
- Rapid determination of the fatigue curve by the thermographic methodInternational Journal of Fatigue, 2002
- Thermographic methodology for rapid determination of the fatigue limit of materials and mechanical componentsInternational Journal of Fatigue, 2000
- Biomechanics of the Hip-JointClinical Orthopaedics and Related Research, 1973