Torsional and cyclic fatigue resistances of glide path preparation instruments: G‐file and PathFile
Abstract
Summary: This study aimed to compare cyclic fatigue and torsional resistances of glide path creating instruments with different tapers and tip sizes. Two sizes (G1 and G2) from G‐File system and three sizes (PathFile #1, #2, and #3) from PathFile system were used for torsional resistance and cyclic fatigue resistance tests (n = 10). The torsional resistance was evaluated at 2‐, 3‐, 4‐, 5‐, and 6‐mm from the file tip by plotting the torsional load changes until fracture by rotational loading of 2 rpm. The cyclic fatigue resistance was compared by measuring the number of cycles to failure. Data were analyzed statistically using one‐way ANOVA and Duncan's post‐hoc comparison. The length of the fractured file fragment was also measured. All fractured fragments were observed under a scanning electron microscope (SEM). Although G‐2 file showed a lower torsional strength than PathFile #3 at 2‐ and 3‐mm levels (p < 0.05), they had similar ultimate strengths at 4‐, 5‐, and 6‐mm levels (p > 0.05). The smaller files of each brand had a significantly higher cyclic fatigue resistance than the bigger ones (p < 0.05). PathFile #1 and #2 had higher fatigue resistances than G‐files (p < 0.05). While G‐1 had a similar fatigue resistance as PathFile #3, G‐2 showed the lowest and PathFile #1 showed the highest resistances among the tested groups (p < 0.05). The SEM examination showed typical appearances of cyclic fatigue and torsional fractures, regardless of the tested levels. Clinicians may consider the instruments' sizes for each clinical case in order to get efficient glide path with minimal risk of fracture. SCANNING 36:500–506, 2014.Keywords
Funding Information
- 2-Year Research Grant of Pusan National University
This publication has 24 references indexed in Scilit:
- Comparison among Manual Instruments and PathFile and Mtwo Rotary Instruments to Create a Glide Path in the Root Canal Preparation of Curved CanalsJournal of Endodontics, 2012
- Comparison of Torsional Stiffness of Nickel-Titanium Rotary Files with Different Geometric CharacteristicsJournal of Endodontics, 2011
- A Retrospective Study of Endodontic Treatment Outcome between Nickel-Titanium Rotary and Stainless Steel Hand Filing TechniquesJournal of Endodontics, 2009
- Use of Nickel-Titanium Rotary PathFile to Create the Glide Path: Comparison With Manual Preflaring in Simulated Root CanalsJournal of Endodontics, 2009
- Effect of Environment on Low-cycle Fatigue of a Nickel–Titanium InstrumentJournal of Endodontics, 2007
- Instrument fracture: mechanisms, removal of fragments, and clinical outcomesEndodontic Topics, 2007
- SEM Observations of Nickel-Titanium Rotary Endodontic Instruments that Fractured During Clinical UseJournal of Endodontics, 2005
- Influence of Manual Preflaring and Torque on the Failure Rate of ProTaper Rotary InstrumentsJournal of Endodontics, 2004
- Analysis of Mechanical Preparations in Extracted Teeth Using ProTaper Rotary Instruments: Value of the Safety QuotientJournal of Endodontics, 2003
- Revised ANSI/ADA specifications no. 28* for root canal files and reamers, type K, and no. 58+ for root canal files, type H (Hedstrom)The Journal of the American Dental Association, 1989