Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients
Authors:
- 7 December 2017
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 32 (5) , 953-964
Abstract
Subcutaneous continuous glucose monitoring (CGM) may have benefits in achieving glycemic control in critically ill patients. The aim of this study was to assess the accuracy and reliability of the FreeStyle Navigator I in critically ill patients and to assess patient related factors influencing the accuracy and reliability. This study is a retrospective analysis of data from a randomized controlled trial conducted in a 20-bed mixed intensive care unit. Analytical accuracy, clinical accuracy and reliability were assessed against arterial blood glucose samples as reference. Assessment was according to recent consensus recommendations with median absolute relative difference (median ARD), Bland–Altman plots, the ISO system accuracy standards (ISO 15197:2013) and Clarke error grid analysis (CEG). We analyzed 2840 paired measurements from 155 critically ill patients. The median ARD of all paired values was 13.3 [6.9–22.1]%. The median ARD was significantly higher in both the hypoglycemic and the hyperglycemic range (32.4 [12.1–53.4]% and 18.7 [10.7–28.3]% respectively, p < 0.001). The Bland–Altman analysis showed a mean bias of − 0.82 mmol/L with a lower limit of agreement (LOA) of − 3.88 mmol/L and an upper LOA of 2.24 mmol/L. A total of 1626 (57.3%) values met the ISO-2013, standards and 1,334 (47%) CGM values were within 12.5% from the reference value. CEG: 71.0% zone A, 25.8% zone B, 0.5% zone C, 2.5% zone D, 0.3% zone E. The median overall real-time data display time was 94.0 ± 14.9% and in 23% of the patients, the sensor measured < 95% of the time. Additionally, data gaps longer than 30 min were found in 48% of the patients. The analytical accuracy of the FreeStyle Navigator I in critically ill patients was suboptimal. Furthermore, the clinical accuracy, did not meet the required standards. The reliability was satisfactory, however, in almost a quarter of the patients the realtime data display was < 95%. The accuracy was considerably and significantly lower in hyper- and hypoglycemic ranges.Keywords
This publication has 48 references indexed in Scilit:
- Feasibility of fully automated closed-loop glucose control using continuous subcutaneous glucose measurements in critical illness: a randomized controlled trialCritical Care, 2013
- Clinical review: Consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adultsCritical Care, 2013
- Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective StudyPLOS ONE, 2013
- The Use of Continuous Glucose Monitoring Combined with Computer-Based eMPC Algorithm for Tight Glucose Control in Cardiosurgical ICUBioMed Research International, 2013
- Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients*Critical Care Medicine, 2011
- Intensive versus Conventional Glucose Control in Critically Ill PatientsNew England Journal of Medicine, 2009
- Management of Hyperglycemia in the Hospital SettingNew England Journal of Medicine, 2006
- Intensive Insulin Therapy in the Medical ICUNew England Journal of Medicine, 2006
- Association Between Hyperglycemia and Increased Hospital Mortality in a Heterogeneous Population of Critically Ill PatientsMayo Clinic Proceedings, 2003
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986