Improved estimation of fructosamine, as a measure of glycated serum protein, with the Technicon RA-1000 analyzer.
Authors:
- F San-GilFSF San-Gil
- G M SchierGSG M Schier
- R G MosesRMR G Moses
- I E GanIGI E Gan
Open Access
- 1 December 1985
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 31 (12) , 2005-2006
Abstract
We present an improved fructosamine assay for use with the Technicon RA-1000 analyzer. Features of this assay include: specimen throughput of 100 per hour, within-batch CV of 1.5%, between-batch CV of 2.4%, standard curve linear up to 24.9 mmol of fructosamine per liter, and good correlation (r = 0.82) with hemoglobin A1.Keywords