In vivo accuracy of image guidance performed using optical tracking and optimized registration
- 1 October 2006
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 105 (4) , 561-567
Abstract
Object: Image guidance systems involving the use of frameless referencing of surgical space to compile volumetric imaging data sets recently have come into widespread use. Few studies have addressed the true intraoperative surgical accuracy (that is, the application accuracy) of these systems except in a subjective manner. Calculated accuracies given by the systems do not necessarily reflect true intraoperative accuracy. Methods: To objectively assess the stereotactic accuracy of a frameless image guidance system using optical spatial referencing, the author analyzed postoperative magnetic resonance (MR) images after placement of depth electrodes for the investigation of epilepsy. Preoperative planning for the treatment of seven patients included implanting skull fiducial screws and obtaining computed tomography/MR fusion images by using ImMerge image fusion software on the StealthStation (Medtronic, Inc.). A total of 42 electrodes were placed. Postoperative volumetric MR images were fused with preoperative study images. The difference between the planned electrode trajectories and targets and the visualized electrodes was measured in stereotactic space. Conclusions: The mean distance between the distal electrode contact and the distal end of the planned trajectory for the 42 targets was 3 ± 1.5 mm. The most common error was in depth. The author’s technique did not involve rigid skull fixation of electrodes because they were subsequently tunneled subcutaneously and later removed at the bedside of the patient. Errors in depth were known to be due to traction at the time of tunneling and not due to stereotactic factors. Correcting for depth along the electrode trajectory, the mean accuracy was found to be 2.4 ± 1 mm.Keywords
This publication has 22 references indexed in Scilit:
- Frameless stereotactic placement of depth electrodes in epilepsy surgeryJournal of Neurosurgery, 2005
- Insertion of depth electrodes with or without subdural grids using frameless stereotactic guidance systems -- technique and outcomeBritish Journal Of Neurosurgery, 2002
- Fiducial Point Placement and the Accuracy of Point-based, Rigid Body RegistrationNeurosurgery, 2001
- Neurophysiological Monitoring for Epilepsy Surgery: The Talairach SEEG MethodStereotactic and Functional Neurosurgery, 2001
- Clinical Evaluation of Multimodality Registration in Frameless StereotaxyComputer Aided Surgery, 1999
- Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studiesJournal of Neurosurgery, 1999
- MRI Distortion and Stereotactic Neurosurgery Using the Cosman-Roberts-Wells and Leksell FramesStereotactic and Functional Neurosurgery, 1996
- MRI confirmation of accuracy of freehand placement of mesial temporal lobe depth electrodes in the investigation of intractable epilepsyBritish Journal Of Neurosurgery, 1996
- Effects of coregistration of MR to CT images on MR stereotactic accuracyJournal of Neurosurgery, 1995
- The Application Accuracy of Stereotactic FramesNeurosurgery, 1994