Measurement of Breast Density with Dual X-ray Absorptiometry: Feasibility

Authors:
  • KKKarla M. Kerlikowske
    Karla M. Kerlikowske
  • RSRebecca Smith-Bindman
    Rebecca Smith-Bindman
  • HGHarry K. Genant
    Harry K. Genant
  • SCSteve R. Cummings
    Steve R. Cummings

Abstract
Dual x-ray absorptiometry (DXA) was used to quantify breast density with a phantom and with cadaveric breasts. With DXA, percentage of fat correlated with percentage of glandular density of the phantom (r > 0.998) and with density at mammography (r(adjusted) = 0.83). DXA precision (SD) was 0.5% without and 1.1% with breast repositioning. DXA devices can be used to accurately and precisely estimate breast tissue density.

This publication has 24 references indexed in Scilit: