Initial anatomic investigations of the I‐gel airway: a novel supraglottic airway without inflatable cuff*
Authors:
- 19 August 2005
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 60 (10) , 1022-1026
Abstract
Summary: The I‐gel airway is a novel supraglottic airway that uses an anatomically designed mask made of a gel‐like thermoplastic elastomer. We studied the positioning and mechanics of this new device in 65 non‐embalmed cadavers with 73 endoscopies (eight had repeat insertion), 16 neck dissections, and six neck radiographs. A full view of the glottis (percentage of glottic opening score 100%) occurred in 44/73 insertions, whereas only 3/73 insertions had epiglottis‐only views. Including the eight repeat insertions with a different size, a glottic opening score of > 50% was obtained in all 65 cadavers. The mean percentage of glottic opening score for the 73 insertions was 82% (95% confidence interval 75–89%). In each of the neck dissections and radiographs the bowl of the device covered the laryngeal inlet. We found that the I‐gel effectively conformed to the perilaryngeal anatomy despite the lack of an inflatable cuff and it consistently achieved proper positioning for supraglottic ventilation.Keywords
This publication has 13 references indexed in Scilit:
- A Proposed Classification and Scoring System for Supraglottic Sealing Airways: A Brief ReviewAnesthesia & Analgesia, 2004
- Bilateral hypoglossal nerve injury following the use of the laryngeal mask airwayAnaesthesia, 2002
- Swelling and Cyanosis of the Tongue Associated with Use of a Laryngeal Mask AirwayAnaesthesia and Intensive Care, 2000
- The Seeing Stylet: a new device for tracheal intubationResuscitation, 2000
- Assessment of laryngeal view: Percentage of glottic opening scorevs Cormack and Lehane gradingCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1999
- Recurrent Laryngeal Nerve Injury Caused by a Laryngeal Mask AirwayAnaesthesia and Intensive Care, 1999
- A Comparison of Pharyngeal Mucosal Pressure and Airway Sealing Pressure with the Laryngeal Mask Airway in Anesthetized Adult PatientsAnesthesia & Analgesia, 1998
- A Comparison of the Disposable Versus the Reusable Laryngeal Mask Airway in Paralyzed Adult PatientsAnesthesia & Analgesia, 1998
- Assessment of Airway Visualization: Validation of the Percentage of Glottic Opening (POGO) ScaleAcademic Emergency Medicine, 1998
- Influence of cuff volume on oropharyngeal leak pressure and fibreoptic position with the laryngeal mask airwayBritish Journal of Anaesthesia, 1998