80-MHz intravascular ultrasound transducer using PMN-PT free-standing film
Authors:
- 10 November 2011
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 58 (11) , 2281-2288
Abstract
[Pb(Mg 1/3 Nb 2/3 )O 3 ] 0.63 [PbTiO 3 ] 0.37 (PMN-PT) free-standing film of comparable piezoelectric properties to bulk material with thickness of 30 μm has been fabricated using a modified precursor coating approach. At 1 kHz, the dielectric permittivity and loss were 4364 and 0.033, respectively. The remnant polarization and coercive field were 28 μC/ cm 2 and 18.43 kV/cm. The electromechanical coupling coefficient k t was measured to be 0.55, which was close to that of bulk PMN-PT single-crystal material. Based on this film, high-frequency (82 MHz) miniature ultrasonic transducers were fabricated with 65% bandwidth and 23 dB insertion loss. Axial and lateral resolutions were determined to be as high as 35 and 176 μ m. In vitro intravascular imaging on healthy rabbit aorta was performed using the thin film transducers. In comparison with a 35-MHz IVUS transducer, the 80-MHz transducer showed superior resolution and contrast with satisfactory penetration depth. The imaging results suggest that PMN-PT free-standing thin film technology is a feasible and efficient way to fabricate very-high-frequency ultrasonic transducers.Keywords
This publication has 20 references indexed in Scilit:
- Piezoelectric films for high frequency ultrasonic transducers in biomedical applicationsProgress in Materials Science, 2011
- Thickness‐Dependent Properties of Relaxor‐PbTiO3 Ferroelectrics for Ultrasonic TransducersAdvanced Functional Materials, 2010
- Self-separated hydrothermal lead zirconate titanate thick films for high frequency transducer applicationsApplied Physics Letters, 2009
- Piezoelectric, ferroelectric Pb(Mg1/3Nb2/3)O3–PbTiO3 thin films with compositions around the morphotropic phase boundary prepared by a sol-gel process of reduced thermal budgetJournal of Materials Research, 2009
- Processing and properties of 0.65Pb(Mg1/3Nb2/3)O3–0.35PbTiO3 thick filmsJournal of the European Ceramic Society, 2008
- Lead zirconate titanate thick film with enhanced electrical properties for high frequency transducer applicationsApplied Physics Letters, 2008
- Self-focused ZnO transducers for ultrasonic biomicroscopyJournal of Applied Physics, 2008
- Pb(Mg1/3Nb2/3)O3–PbTiO3 thick films from mechanochemically synthesized powderJournal of the European Ceramic Society, 2007
- Correlation between high frequency intravascular ultrasound and histomorphology in human coronary arteriesHeart, 2001
- Advances in ultrasound biomicroscopyUltrasound in Medicine & Biology, 2000