Identification of Hemodynamically Optimal Coronary Stent Designs Based on Vessel Caliber
- 1 July 2012
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 59 (7) , 1992-2002
Abstract
Coronary stent design influences local patterns of wall shear stress (WSS) that are associated with neointimal growth, restenosis, and the endothelialization of stent struts. The number of circumferentially repeating crowns N C for a given stent de- sign is often modified depending on the target vessel caliber, but the hemodynamic implications of altering N C have not previously been studied. In this investigation, we analyzed the relationship between vessel diameter and the hemodynamically optimal N C using a derivative-free optimization algorithm coupled with computational fluid dynamics. The algorithm computed the optimal vessel diameter, defined as minimizing the area of stent-induced low WSS, for various configurations (i.e., N C ) of a generic slotted-tube design and designs that resemble commercially available stents. Stents were modeled in idealized coronary arteries with a vessel diameter that was allowed to vary between 2 and 5 mm. The results indicate that the optimal vessel diameter increases for stent configurations with greater N C , and the designs of current commercial stents incorporate a greater N C than hemodynamically optimal stent designs. This finding suggests that reducing the N C of current stents may improve the hemodynamic environment within stented arteries and reduce the likelihood of excessive neointimal growth and thrombus formation.Keywords
This publication has 46 references indexed in Scilit:
- Optimization of Cardiovascular Stent Design Using Computational Fluid DynamicsJournal of Biomechanical Engineering, 2012
- Multiobjective design optimisation of coronary stentsBiomaterials, 2011
- Studies on Design Optimization of Coronary StentsJournal of Medical Devices, 2008
- Comparison of Effectiveness of Bare Metal Stents Versus Drug-Eluting Stents in Large (≥3.5 mm) Coronary ArteriesThe American Journal of Cardiology, 2007
- Trailing-edge noise reduction using derivative-free optimization and large-eddy simulationJournal of Fluid Mechanics, 2007
- Strut Position, Blood Flow, and Drug DepositionCirculation, 2005
- Is overdilatation of 3.0 mm sirolimus-eluting stent associated with a higher restenosis rate?Catheterization and Cardiovascular Interventions, 2005
- A survey of stent designsMinimally Invasive Therapy & Allied Technologies, 2002
- Acute and long‐term outcomes of stenting in coronary vessel > 3.0 mm, 3.0–2.5 mm, and < 2.5 mmCatheterization and Cardiovascular Interventions, 2001
- Compensatory Enlargement of Human Atherosclerotic Coronary ArteriesThe New England journal of medicine, 1987