Method of recording bioelectrical signals using a capacitive coupling
Open Access
- 27 November 2017
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 929 (1) , 012016
Abstract
In this article a technique for the bioelectrical signals acquisition by means of the capacitive sensors is described. A feedback loop for the ultra-high impedance biasing of the input instrumentation amplifier, which provides receiving of the electrical cardiac signal (ECS) through a capacitive coupling, is proposed. The mains 50/60 Hz noise is suppressed by a narrow-band stop filter with an independent notch frequency and quality factor tuning. Filter output is attached to a ΣΔ analog-to-digital converter (ADC), which acquires the filtered signal with a 24–bit resolution. Signal processing board is connected through universal serial bus interface to a personal computer, where ECS in a digital form is recorded and processed.Keywords
This publication has 4 references indexed in Scilit:
- A method of cardiac signal registration through a capacitive coupling between sensors and patient's skin2016
- Wireless and Non-contact ECG Measurement System – the “Aachen SmartChair”Acta Polytechnica, 2007
- Electric potential probes - new directions in the remote sensing of the human bodyMeasurement Science and Technology, 2001
- An ultra-low-noise electrical-potential probe for human-body scanningMeasurement Science and Technology, 2000