Biomechanical study of olecranon plate fixation in comminuted osteoporotic fracture model: locking compression plates versus dynamic compression plates
- 1 March 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Current Orthopaedic Practice
- Vol. 29 (2) , 167-172
Abstract
This cadaveric biomechanical study compared locking compression plates (LCP) to dynamic compression plates (DCP) in a comminuted olecranon fracture using cyclic loading to measure fragment displacement and maximal failure load. Eight paired human cadavers were used. We confirmed all specimens were osteopenic using DEXA. One paired half of the specimens randomly received LCP fixation and the other DCP. Implants were secured to the intact ulna prior to creating an osteotomy. Two vertical osteotomies were created to simulate a fracture gap of 5 mm. The specimens were loaded and placed on the MTS machine. Saw-toothed cyclic loading was applied and force/gap displacement measured. Failure was defined as fracture gap greater than 2 mm or catastrophic failure of the bone/construct. Biomechanical parameters were compared between two groups. The average force for the DCP was 22.12 N compared to 33.40 N (PPPP<0.05). This implies that the LCP maintained the reduction better than the DCP. No differences in force/gap were noted when comparing gender, right versus left forearm, or osteoporotic versus osteopenic bone. There was no difference when failure occurred between the DCP and LCP groups. Our study found that LCP was able to withstand higher force than DCP, and there was less displacement of the fracture with biomechanical testing in the LCP compared to the DCP. Both could be appropriate surgical techniques for fixation of comminuted olecranon fractures. When treating comminuted and osteopenic fractures, LCP are advantageous. Further studies would be needed to compare these two plates clinically.Keywords
This publication has 28 references indexed in Scilit:
- An Off-Loading Triceps Suture for Augmentation of Plate Fixation in Comminuted Osteoporotic Fractures of the OlecranonJournal of Orthopaedic Trauma, 2012
- Results of open reduction and plate osteosynthesis in comminuted fracture of the olecranonJournal of Shoulder and Elbow Surgery, 2011
- Biomechanical evaluation of fixation of comminuted olecranon fractures: one-third tubular versus locking compression platingArchives of orthopaedic and trauma surgery, 2009
- Fresh‐frozen vs. embalmed bone: is it possible to use formalin‐fixed human bone for biomechanical experiments on implants?Clinical Oral Implants Research, 2009
- A Mechanical Comparison of the Locking Compression Plate (LCP) and the Low Contact-Dynamic Compression Plate (DCP) in an Osteoporotic Bone ModelJournal of Orthopaedic Trauma, 2008
- Congruent Elbow Plate Fixation of Olecranon FracturesJournal of Orthopaedic Trauma, 2007
- Role of Dual-Energy X-Ray Absorptiometry in the Diagnosis and Treatment of OsteoporosisJournal of Clinical Densitometry, 2007
- Comminuted olecranon fractures: A comparison of plating methodsJournal of Shoulder and Elbow Surgery, 2006
- Plate fixation of comminuted olecranon fractures: An in vitro biomechanical studyJournal of Shoulder and Elbow Surgery, 1996
- Olecranon FracturesClinical Orthopaedics and Related Research, 1992