Hemoximetry as the “Gold Standard”? Error Assessment Based on Differences Among Identical Blood Gas Analyzer Devices of Five Manufacturers
Authors:
- Hartmut GehringHGHartmut Gehring
- Lutz DuembgenLDLutz Duembgen
- Mareike PeterleinMPMareike Peterlein
- Söhnke HagelbergSHSöhnke Hagelberg
- Leif DibbeltLDLeif Dibbelt
- 1 December 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 105 (6) , S24-S30
Abstract
The calibration and testing procedures of a pulse oximeter with arterial blood samples from healthy subjects are based on reference values from the hemoximeter. There are no tests to identify the accuracy of the reference devices. Because of this limitation and since the true values of oxygen saturation (sO2 in %) in blood samples were not known, we used the differences between two identical devices, A and B, for error assessment. Two identical devices, A and B, from five leading manufacturers were investigated. Seventy-two arterial blood samples from 12 healthy volunteers at three different levels of saturation between 100% and 70% sO2 were randomly evaluated by the test systems. The observed differences (Delta) between Devices A and B, as a measure for the error of the hemoximeters, increased significantly with all manufacturers from level 97 (Deltamin, -0.9%; Deltamax, 2.6%) to 85 (Deltamin, -2.4%; Deltamax, 4.3), this effect was even stronger between levels 97 and 75 (Deltamin, -4.6%; Deltamax, 4.3%). A variance proportion analysis revealed the concentration of the reduced hemoglobin as the main error source for sO2 measurements. Independent from the sO2 levels there were also significant differences for the carboxy hemoglobin concentration in the range of 0%-4% and for the methemoglobin concentration in the range of 0%-1%. The variance of sO2 measurements between identical devices increased significantly when saturation decreased from the normal level of 97% to the hypoxemic levels of 85% and 75%.Keywords
This publication has 10 references indexed in Scilit:
- Measurement of Carboxyhemoglobin and Methemoglobin by Pulse OximetryAnesthesiology, 2006
- Monitoring Carbon Dioxide Tension and Arterial Oxygen Saturation by a Single Earlobe Sensor in Patients With Critical Illness or Sleep ApneaChest, 2005
- Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low SaturationAnesthesiology, 2005
- The desaturation response time of finger pulse oximeters during mild hypothermia*Anaesthesia, 2004
- Reflectance Pulse Oximetry from Core Body in Neonates and Infants: Comparison to Arterial Blood Oxygen Saturation and to Transmission Pulse OximetryJournal of Perinatology, 2004
- Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsisIntensive Care Medicine, 2004
- Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrationsActa Anaesthesiologica Scandinavica, 2002
- Accuracy of pulse oximetry in the intensive care unitIntensive Care Medicine, 2001
- Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery RoomAnesthesia & Analgesia, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986