Theory and application of array coils in MR spectroscopy
- 4 December 1997
- journal article
- research article
- Published by Wiley in NMR in Biomedicine
- Vol. 10 (8) , 394-410
Abstract
The theory and application of array coils are reviewed in the context of phased array spectroscopy. The optimization of the signal‐to‐noise ratio from an array of coils is developed by considering the efficiency of a phased array transmit coil. This approach avoids the need to consider noise correlation, and should be useful in future considerations of transmit phased array coils for MR spectroscopy. Methods to characterize array coil performance, including fields and coupling are briefly summarized, along with methods to minimize the effects of mutual inductance. The signal‐to‐noise advantages of array coils over single coils are examined for both planar and cylindrical arrays. Numerical simulations of planar arrays of 2×2, 4×4 and 8×8 elements and constant overall dimension are compared to a single coil of the same size. The results demonstrate a significant improvement in sensitivity near the array coil. Although the benefits of the array decrease as a function of distance from the array, the array sensitivity never drops below that of a single coil with the same overall dimensions, or that of a single element of the array. Similar results are obtained for a sixteen element cylindrical array, which is compared to a standard quadrature birdcage coil using both computational methods and phantom measurements. The phased array techniques reviewed are demonstrated with proton spectroscopic images of the brain. © 1997 John Wiley & Sons, Ltd.Keywords
This publication has 58 references indexed in Scilit:
- Partial volume effects in volume localized phased-array proton spectroscopy of the temporal lobeJournal of Magnetic Resonance Imaging, 1995
- Phased Array Coils for Upper Extremity MRAMagnetic Resonance in Medicine, 1995
- Calculation of the signal-to-noise ratio for simple surface coils and arrays of coils [magnetic resonance imaging]IEEE Transactions on Biomedical Engineering, 1995
- High frequency volume coils for clinical NMR imaging and spectroscopyMagnetic Resonance in Medicine, 1994
- MR flow measurement using RF phase gradients in receiver coil arraysIEEE Transactions on Biomedical Engineering, 1993
- Letter to the editorMagnetic Resonance in Medicine, 1992
- Noise correlation in multicoil receiver systemsMagnetic Resonance in Medicine, 1992
- An analytical model for the design of RF resonators for MR body imagingMagnetic Resonance in Medicine, 1991
- A decoupled coil detector array for fast image acquisition in magnetic resonance imagingMedical Physics, 1991
- Influence of skin depth on NMR coil impedancePhysics in Medicine & Biology, 1988