A spatially resolving x-ray crystal spectrometer for measurement of ion-temperature and rotation-velocity profiles on the Alcator C-Mod tokamak
Authors:
- 1 October 2008
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 79 (10) , 10E320
Abstract
A new spatially resolving x-ray crystal spectrometer capable of measuring continuous spatial profiles of high resolution spectra of He-like and H-like lines with good spatial and temporal resolutions has been installed on the Alcator -Mod tokamak. Two spherically bent crystals image the spectra onto four two-dimensional Pilatus II pixel detectors. Tomographic inversion enables inference of local line emissivity, ion temperature , and toroidal plasma rotation velocity from the line Doppler widths and shifts. The data analysis techniques, and profiles, analysis of fusion-neutron background, and predictions of performance on other tokamaks, including ITER, will be presented.