Iterative RF pulse design for multidimensional, small‐tip‐angle selective excitation
- 9 September 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 54 (4) , 908-917
Abstract
The excitation k‐space perspective on small‐tip‐angle selective excitation has facilitated RF pulse designs in a range of MR applications. In this paper, k‐space‐based design of multidimensional RF pulses is formulated as a quadratic optimization problem, and solved efficiently by the iterative conjugate‐gradient (CG) algorithm. Compared to conventional design approaches, such as the conjugate‐phase (CP) method, the new design approach is beneficial in several regards. It generally produces more accurate excitation patterns. The improvement is particularly significant when k‐space is undersampled, and it can potentially shorten pulse lengths. A prominent improvement in accuracy is also observed when large off‐resonance gradients are present. A further boost in excitation accuracy can be accomplished in regions of interest (ROIs) if they are specified together with “don't‐care” regions. The density compensation function (DCF) is no longer required. In addition, regularization techniques allow control over integrated and peak pulse power. Magn Reson Med, 2005.Keywords
This publication has 30 references indexed in Scilit:
- Small tip angle three‐dimensional tailored radiofrequency slab‐select pulse for reduced B1 inhomogeneity at 3 TMagnetic Resonance in Medicine, 2005
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imagingMagnetic Resonance in Medicine, 2002
- Assessment of Reactive Hyperaemia Using Real Time Zonal Echo-Planar Flow ImagingJournal of Cardiovascular Magnetic Resonance, 2002
- An inverse problem approach to the correction of distortion in EPI imagesIEEE Transactions on Medical Imaging, 2000
- Iterative reconstruction of single-shot spiral MRI with off resonanceIEEE Transactions on Medical Imaging, 1999
- A linear class of large-tip-angle selective excitation pulsesJournal of Magnetic Resonance (1969), 1989
- A k-space analysis of small-tip-angle excitationJournal of Magnetic Resonance (1969), 1989
- Optimization of two-dimensional spatially selective NMR pulses by simulated annealingJournal of Magnetic Resonance (1969), 1988
- Computer-optimized narrowband pulses for multislice imagingJournal of Magnetic Resonance (1969), 1987