A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION Study
Open Access
- 7 May 2019
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 21 (5) , 231-237
Abstract
Background: A prior study (PRECISE II) demonstrated that an implantable continuous glucose monitoring (CGM) system (Eversense ® CGM System) provided accurate glucose readings through the 90-day sensor life with a favorable safety profile in participants with type 1 or type 2 diabetes (T1D, T2D). This study was performed to further characterize the accuracy of the system. Methods: PRECISION was a prospective multicenter study that evaluated the accuracy and safety of Eversense among adults with T1D or T2D through 90 days (NCT02647905). Accuracy measures included percentage system agreement and mean absolute relative difference (MARD) between Eversense and Yellow Springs Instrument reference measurements from 40 to 400 mg/dL. The primary safety endpoint was incidence of device-related or sensor insertion/removal procedure-related serious adverse events (SAEs) through 90 days. An updated glucose calculation algorithm was also applied to the sensor data from the PRECISE II study to evaluate consistency of accuracy results. Results: Thirty-five participants received the CGM system. Eighty-five percent of CGM values were within 15/15% of reference and the MARD value against reference was 9.6% (95% confidence interval [CI]: 8.9–10.4). All sensors were functional through day 90. No device- or procedure-related SAEs occurred. Application of the updated algorithm to PRECISE II sensor data resulted in 87% of readings within 15/15% of reference and an MARD value against reference of 8.5% (95% CI: 8.0%–9.1%). Conclusions: PRECISION corroborated prior accuracy and safety findings of the Eversense CGM System through the 90-day sensor life. The updated algorithm improved accuracy of measurements in PRECISE II.Keywords
This publication has 11 references indexed in Scilit:
- A Prospective Multicenter Evaluation of the Accuracy of a Novel Implanted Continuous Glucose Sensor: PRECISE IIDiabetes Technology & Therapeutics, 2018
- Simultaneous assay of cortisol and dexamethasone improved diagnostic accuracy of the dexamethasone suppression testEuropean Journal of Endocrinology, 2017
- Effect of Continuous Glucose Monitoring on Glycemic Control in Adults With Type 1 Diabetes Using Insulin InjectionsJAMA, 2017
- Continuous Glucose Monitoring vs Conventional Therapy for Glycemic Control in Adults With Type 1 Diabetes Treated With Multiple Daily Insulin InjectionsJAMA, 2017
- Accuracy and Longevity of an Implantable Continuous Glucose Sensor in the PRECISE Study: A 180-Day, Prospective, Multicenter, Pivotal TrialDiabetes Care, 2016
- Time-dependent effects of dexamethasone plasma concentrations on glucocorticoid receptor challenge testsPsychoneuroendocrinology, 2016
- Increased in vivo stability and functional lifetime of an implantable glucose sensor through platinum catalysisJournal of Biomedical Materials Research Part A, 2012
- The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trialDiabetologia, 2012
- Short- and Long-Term Effects of Real-Time Continuous Glucose Monitoring in Patients With Type 2 DiabetesDiabetes Care, 2011
- Validation of Measures of Satisfaction with and Impact of Continuous and Conventional Glucose MonitoringDiabetes Technology & Therapeutics, 2010