Comparative evaluation of BACTEC MGIT 960 with BACTEC 9000 MB and LJ for isolation of mycobacteria in The Gambia

Authors:

Abstract
Background: The BACTEC MGIT 960 was evaluated and compared with BACTEC 9000 MB and Lowenstein-Jensen medium for recovery rate of mycobacteria, time to detection, and contamination rate. Methodology: 147 sputum samples obtained from patients with suspicion of tuberculosis were processed and inoculated into BACTEC MGIT 960, BACTEC 9000 MB and Lowenstein-Jensen medium using standardized procedures. Results: BACTEC MGIT 960 detected 57.1%; BACTEC 9000 MB detected 57.8%; and LJ medium detected 43.5% specimens with Mycobacterium tuberculosis complex (MTBC). BACTEC MGIT 960 had the shortest mean number of days (10.3) to detection, followed by BACTEC 9000 MB (13.2) and LJ medium (26.1). Sign rank test showed all three methods had significant difference in days to detection (each P