Recent Advances in Electret Ion Chamber Technology for Radiation Measurements

Authors:
  • PKP. Kotrappa
    P. Kotrappa
  • JDJ.C. Dempsey
    J.C. Dempsey
  • LSL.R. Steiff
    L.R. Steiff

Abstract
The use of electret ion chamber technology (EIC) for radiation measurement has been investigated for many years, but its practical use is of recent origin. A family of EIC-based devices has been developed and the technology has found a new niche in the general field of radiation dosimetry. The most significant advance is the commercial availability of the EIC system, consisting of stable electrets and chambers made of electrically conducting plastic. These are primarily used as passive devices wherein an electret (a charged disc of Teflon) is loaded into the chamber to form an EIC. An EIC functions as an integrating ionisation chamber with the electret serving as both a source of electrical potential to collect the ions and as a quantitative sensor. The voltage drop of the electret over a time interval is a measure of total ionisation in that time interval in the volume of the chamber. The EIC is most widely used for radon and radon related measurements. Other measurement applications include, beta and gamma dosimetry, airborne tritium and tritiated water vapour, low level alpha and beta contamination, and neutrons. The status of the technology and the possible future applications are presented.