Design and Characteristics of 3D Magnetically Steerable Guidewire System for Minimally Invasive Surgery
- 1 February 2022
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Robotics and Automation Letters
- Vol. 7 (2) , 4040-4046
Abstract
Endovascular techniques have been increasingly adapted to medical applications as a minimally-invasive treatment approach to diagnose and treat various vascular diseases. Guidewires are the basic devices in endovascular surgery. The conventional practice of endovascular procedures has a number of drawbacks such as prolonged exposure to high doses of X-ray radiation, long operation time and undesired damage by surgeon’s unskilled operation. In this letter, we proposed a magnetically steerable guidewire and designed a three-dimensional magnetically actuating system. The proposed magnetically steerable guidewire consists of a commercial guidewire connecting to a soft actuator made of Ecoflex and magnetic powder. The actuating system contains two symmetrical parts, each consisting of three linear motors. It has great workspace of 700 mm × 500 mm × 500 mm, which can help the proposed guidewire to achieve a wide range of deflection angle. The relationship between the proposed guidewire’s deflection angle and the magnitude of the external magnetic field are explored, as well as the direction of the magnetic field. The proposed guidewire is flexible enough to make sharp turns up to 180 ${}^{\circ }$ in three dimensional workspace. The guidewire has achieved navigating through a set of rings and in a complex and constrained vascular phantom by programmed actuating magnetic fields.Keywords
Funding Information
- National Natural Science Foundation of China (62022087)
- Youth Innovation Promotion Association of CAS
- Special Support Project for Outstanding Young Scholars of Guangdong Province (2019TQ05X933)
- CAS-Croucher Funding Scheme for Joint Laboratories (172644KYSB20200044)
- Croucher Foundation (CAS20403)
- Shenzhen Key Basic Research projection (JCYJ20200109114827177)
- SIAT-CUHK Joint Laboratory of Robotics and Intelligent Systems
This publication has 30 references indexed in Scilit:
- Bioinspiration & Biomimetics2020
- Small-scale soft-bodied robot with multimodal locomotionNature, 2018
- Current applications and future perspectives of robotics in cerebrovascular and endovascular neurosurgeryJournal of NeuroInterventional Surgery, 2017
- Feasibility study on magnetically steerable guidewire device for percutaneous coronary interventionInternational Journal of Control, Automation, and Systems, 2016
- Shape-programmable magnetic soft matterProceedings of the National Academy of Sciences of the United States of America, 2016
- Robotic technology in cardiovascular medicineNature Reviews Cardiology, 2014
- Current and Emerging Robot-Assisted Endovascular Catheterization Technologies: A ReviewAnnals of Biomedical Engineering, 2013
- Design and Kinematic Modeling of Constant Curvature Continuum Robots: A ReviewThe International Journal of Robotics Research, 2010
- Steerable catheters in minimally invasive vascular surgeryThe International Journal of Medical Robotics and Computer Assisted Surgery, 2009
- Accurate and Reproducible Target Navigation with the Stereotaxis Niobe® Magnetic Navigation SystemJournal of Cardiovascular Electrophysiology, 2006