Effects of voxel size and iterative reconstruction parameters on the spatial resolution of SPECT/CT
Open Access
- 1 September 2011
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 12 (4) , 210-220
Abstract
The purpose of this study was to evaluate the effects of voxel size and iterative reconstruction parameters on the radial and tangential resolution forSPECT as a function of radial distance from isocenter. SPECT/CT scans of eight coplanar point sources of size smaller thancontaining high concentrationsolution were acquired on a SPECT/CT system with 5/8 inch NaI(Tl) detector and low‐energy, high‐resolution collimator. The tomographic projection images were acquired in step‐and‐shoot mode for 360 views over 360° with 250,000 counts per view, a zoom of 2.67, and an image matrix ofpixels that resulted in aSPECT voxel size over 230 mm field‐of‐view. The projection images were also rebinned to image matrices ofandto yield SPECT voxel sizes ofand, respectively. The SPECT/CT datasets were reconstructed using the vendor‐supplied iterative reconstruction software that incorporated collimator‐specific resolution recovery, CT‐based attenuation correction, and dual‐energy window‐based scatter correction using different combinations of iterations and subsets. SPECT spatial resolution was estimated as the full width at half maximum of the radial and tangential profiles through the center of each point source in reconstructed SPECT images. Both radial and tangential resolution improved with higher iterations and subsets, and with smaller voxel sizes. Both radial and tangential resolution also improved with radial distance further away from isocenter. The magnitude of variation decreased for smaller voxel sizes and for higher number of iterations and subsets. Tangential resolution was found not to be equal to the radial resolution, and the nature of the anisotropy depended on the distribution of the radionuclide and on the reconstruction parameters used. The tangential resolution converged faster than the radial resolution, with higher iterations and subsets. SPECT resolution was isotropic and independent of radial distance when reconstructed using filtered back‐projection. SPECT spatial resolution and therefore quantification of SPECT uptake via partial‐volume correction in clinical images were found to depend on the nature of activity distribution within the SPECT field‐of‐view and on the specific choice of iterative reconstruction parameters.PACS number: 87.57.uh, 87.57.cf, 87.87.nfKeywords
This publication has 9 references indexed in Scilit:
- A multi-center phantom study comparing image resolution from three state-of-the-art SPECT-CT systemsJournal of Nuclear Cardiology, 2009
- Towards quantitative SPECT: Error estimation of SPECT OSEM with 3D resolution recovery, attenuation correction and scatter correction2008
- Analysis of the dependence of PET/CT quantification on iterative reconstruction parameters2007
- Collimator-Detector Response Compensation in SPECT2006
- Iterative Reconstruction Methods2006
- Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: phantom studies2004
- Quantitation of SPECT performance: Report of Task Group 4, Nuclear Medicine CommitteeMedical Physics, 1995
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- The convergence of object dependent resolution in maximum likelihood based tomographic image reconstructionPhysics in Medicine & Biology, 1993