Online Calibration of Glucose Sensors From the Measured Current by a Time-Varying Calibration Function and Bayesian Priors
- 24 April 2015
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 63 (8) , 1631-1641
Abstract
Goal: Minimally invasive continuous glucose monitoring (CGM) sensors measure in the subcutis a current signal, which is converted into interstitial glucose (IG) concentration by a calibration process periodically updated using fingerstick blood glucose (BG) references. Though important in diabetes management, CGM sensors still suffer from accuracy problems. Here, we propose a new online calibration method improving accuracy of CGM glucose profiles with respect to manufacturer calibration. Method: The proposed method fits CGM current signal against the BG references collected twice a day for calibration purposes, by a time-varying calibration function whose parameters are identified in a Bayesian framework using a priori second-order statistical knowledge. The distortion introduced by BG-to-IG kinetics is compensated before parameter identification via nonparametric deconvolution. Results: The method was tested on a database where 108 CGM signals were collected for 7 days by the Dexcom G4 Platinum sensor. Results show the new method drives to a statistically significant accuracy improvement as measured by three commonly used metrics: mean absolute relative difference reduced from 12.73% to 11.47%; percentage of accurate glucose estimates increased from 82.00% to 89.19%; and percentage of values falling in the “A” zone of the Clark error grid increased from 82.22% to 88.86%. Conclusion: The new calibration method significantly improves CGM glucose profiles accuracy with respect to manufacturer calibration. Significance: The proposed algorithm provides a real-time improvement of CGM accuracy, which can be crucial in several CGM-based applications, including the artificial pancreas, thus providing a potential great impact in the diabetes technology research community.Keywords
This publication has 34 references indexed in Scilit:
- Modeling the Glucose Sensor ErrorIEEE Transactions on Biomedical Engineering, 2014
- Dexcom G4AP: An Advanced Continuous Glucose Monitor for the Artificial PancreasJournal of Diabetes Science and Technology, 2013
- Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation is Part of the AssessmentJournal of Diabetes Science and Technology, 2013
- Metabolic Biofouling of Glucose Sensors in Vivo: Role of Tissue MicrohemorrhagesJournal of Diabetes Science and Technology, 2011
- New Features and Performance of a Next-Generation SEVEN-Day Continuous Glucose Monitoring System with Short Lag TimeDiabetes Technology & Therapeutics, 2009
- The Chemistry of Commercial Continuous Glucose MonitorsDiabetes Technology & Therapeutics, 2009
- How to Assess and Compare the Accuracy of Continuous Glucose Monitors?Diabetes Technology & Therapeutics, 2008
- The Original Clarke Error Grid Analysis (EGA)Diabetes Technology & Therapeutics, 2005
- Continuous Glucose Monitoring with Glucose Sensors: Calibration and Assessment CriteriaDiabetes Technology & Therapeutics, 2003
- Nonparametric input estimation in physiological systems: Problems, methods, and case studiesAutomatica, 1997