Radiographic Predictors of Lead Conductor Fracture
Authors:
- Andrew D. KrahnAKAndrew D. Krahn
- Josée MorissetteJMJosée Morissette
- Ryan LahmRLRyan Lahm
- Tarek HaddadTHTarek Haddad
- Walt W. BaxterWBWalt W. Baxter
- Rick McVenesRMRick McVenes
- Eugene CrystalECEugene Crystal
- Félix Ayala-ParedesFAFélix Ayala-Paredes
- Doug CameronDCDoug Cameron
- Atul VermaAVAtul Verma
- Christopher S. SimpsonCSChristopher S. Simpson
- Derek V. ExnerDEDerek V. Exner
Open Access
- 1 December 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 7 (6) , 1070-1077
Abstract
Background—Lead fracture is a limiting factor in high voltage lead durability. Fractures noted with the Medtronic Fidelis leads provide an opportunity to examine factors captured on implant chest x-ray that correlate with risk for lead conductor fracture. We evaluated contributory factors in a large population of fractures. Methods and Results—We conducted a retrospective case–control study at 8 Canadian centers that routinely capture anterior posterior and lateral chest x-rays within 2 weeks of implant. Cases were patients that experienced confirmed Medtronic Fidelis 6949 lead fracture based on standard definitions, matched one-to-one to controls for date of implant, sex, and age with normally functioning Fidelis leads from the same center. Select chart data and x-rays were collected for all patients. Radiographic measurements by ≥2 individuals per case/control were blinded to patient status. The data were analyzed using a time to failure multivariable Cox proportional hazards model with stratification for each matched pair. X-ray pairs from 111 fracture patients were compared with 111 controls (age 61.5±12.8 years, 75% male, 221 model 6949 leads). Six parameters included in the statistical analysis were significantly associated with risk of fracture, including slack/tortuosity measures, pulse generator and superior vena cava coil location, and angle of lead exit from the pocket. Conclusions—Pocket, intravascular and intracardiac lead characteristics on x-ray correlate with risk of lead conductor fracture. These observations may be useful to direct implant technique to optimize lead durability. Validation in larger populations and other lead models may inform the application of these results.Keywords
This publication has 28 references indexed in Scilit:
- Transvenous extraction of implantable cardioverter-defibrillator leads under advisory—A comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leadsHeart Rhythm, 2013
- Understanding lead dysfunction: Managing lead failure griefHeart Rhythm, 2013
- The Canadian experience with Durata and Riata ST Optim defibrillator leads: A report from the Canadian Heart Rhythm Society Device CommitteeHeart Rhythm, 2013
- Lead electrical parameters may not predict integrity of the Sprint Fidelis ICD leadHeart Rhythm, 2012
- Clinical Predictors of Fidelis Lead FailureCirculation, 2012
- Complications Associated With Revision of Sprint Fidelis LeadsCirculation, 2010
- Risk Factors for Implantable Defibrillator Lead Fracture in a Recalled and a Nonrecalled LeadJournal of Cardiovascular Electrophysiology, 2010
- Recommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and GuidelinesHeart Rhythm, 2009
- Utilization of a national network for rapid response to the Medtronic Fidelis lead advisory: The Canadian Heart Rhythm Society Device Advisory CommitteeHeart Rhythm, 2009
- Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator leadHeart Rhythm, 2007