Comparing the Accuracy of 2 Point-of-Care Lipid Testing Devices
- 16 June 2016
- journal article
- research article
- Published by SAGE Publications in Journal of Pharmacy Practice
- Vol. 30 (5) , 490-497
Abstract
Background: Device manufacturers have improved technology since studies were last published, thus warranting an updated analysis. Objective: Two point-of-care (POC) cholesterol testing devices were directly compared to a venous sample to determine device accuracy. Methods: Institutional review board (IRB)–approved study collected finger-stick blood samples analyzed by Cholestech LDX (Cholestech Corporation, Hayward, California) and CardioChek Plus (Polymer Technology Systems Inc, Indianapolis, Indiana) devices and compared to venous blood for 30 study participants. Statistical analyses were completed using StatisPro. Intraclass correlation coefficients were generated, and the average difference expected to be within the industry standards of total cholesterol (TC; ±10%), high-density lipoprotein (HDL) cholesterol (±12%), and triglycerides (TG; ±15%). Results: The POC devices produced clinically equivalent values when compared to the same patients’ samples analyzed in a reference laboratory. The average difference calculated from the actual individual paired percentage bias with the Integra analyzer: venous—TC −3.8%, HDL −6.9%, TG −1.8%; CardioChek—TC −7.8%, HDL −6.2%, TG 5.1%; and Cholestech—TC 0.5%, HDL −4.5%, TG −3.3%. The average of the actual paired percentage bias with the Roche Cobas analyzer: CardioChek—TC −4.2%, HDL 0.8%, TG 7.0% and Cholestech—TC 4.6%, HDL 2.6%, TG −1.6%. Conclusion: Both screening devices operated within industry accuracy standards.Keywords
This publication has 6 references indexed in Scilit:
- Treatment of Blood Cholesterol to Reduce Atherosclerotic Cardiovascular Disease Risk in Adults: Synopsis of the 2013 American College of Cardiology/American Heart Association Cholesterol GuidelineAnnals of Internal Medicine, 2014
- Point-of-care screenings at the University of Minnesota: Mechanism for civic engagementJournal of the American Pharmacists Association, 2014
- Comparison of Two Point-of-Care Lipid Analyzers for Use in Global Cardiovascular Risk AssessmentsAnnals of Pharmacotherapy, 2008
- PO13-344 CARDIOCHEK PA CHOLESTEROL-TESTING DEVICE COMPARED TO DIAGNOSTIC LABORATORY METHODAtherosclerosis Supplements, 2007
- Validation of a point of care lipid analyser using a hospital based reference laboratoryIrish Journal of Medical Science (1971 -), 2006
- Cost of a Pharmacist‐Directed Intervention to Increase Treatment of HypercholesterolemiaPharmacotherapy: The Journal of Human Pharmacology and Drug Therapy, 2004