Alterations in femoral strain following hip resurfacing and total hip replacement
Authors:
- 1 January 2009
- journal article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 91-B (1) , 124-130
Abstract
Bone surface strains were measured in cadaver femora during loading prior to and after resurfacing of the hip and total hip replacement using an uncemented, tapered femoral component. In vitro loading simulated the single-leg stance phase during walking. Strains were measured on the medial and the lateral sides of the proximal aspect and the mid-diaphysis of the femur. Bone surface strains following femoral resurfacing were similar to those in the native femur, except for proximal shear strains, which were significantly less than those in the native femur. Proximomedial strains following total hip replacement were significantly less than those in the native and the resurfaced femur.These results are consistent with previous clinical evidence of bone loss after total hip replacement, and provide support for claims of bone preservation after resurfacing arthroplasty of the hip.Keywords
This publication has 41 references indexed in Scilit:
- Femoral neck fracture following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2008
- Birmingham hip resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2008
- The five-year results of the Birmingham Hip Resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2008
- New Femoral DesignsClinical Orthopaedics and Related Research, 2006
- Biomechanics of the Birmingham hip resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2006
- To Resurface or Replace the Hip in the Under 65-Year-OldThe Annals of The Royal College of Surgeons of England, 2006
- Birmingham hip resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2005
- Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1998
- Metal on Metal Surface Replacement of the HipClinical Orthopaedics and Related Research, 1996
- Fit of the uncemented femoral component and the use of cement influence the strain transfer to the femoral cortexJournal of Orthopaedic Research, 1994