Scheimpflug-Based Tomography and Biomechanical Assessment in Pressure-Induced Stromal Keratopathy

Authors:

Abstract
Report Fernando Faria-Correia, MD; Isaac Ramos, MD; Bruno Valbon, MD; Allan Luz, MD; Cynthia J. Roberts, PhD; Renato Ambrósio, MD, PhD To report the tomographic and biomechanical findings before and after treatment of a case of pressure-induced stromal keratopathy (PISK), which was misdiagnosed as diffuse lamellar keratitis (DLK). A case report of a referred patient with supposed diagnosis of DLK after LASIK in the right eye. Scheimpflug-based corneal tomography and biomechanical assessment were provided by the Pentacam HR and CorVis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany). A layer of corneal opacity beneath the flap with a presumably fluid-filled interface area was observed on slit-lamp biomicroscopy. Scheimpflug image from Pentacam revealed a hyperreflective area underneath the flap interface. Goldmann applanation tonometry was 12 mm Hg, whereas CorVis intraocular pressure was 53.5 mm Hg with deformation amplitude of 0.42 mm. Two days after starting oral and topical ocular hypotensive therapy, CorVis intraocular pressure was 14 mm Hg and deformation amplitude was 1.02 mm. Ocular hypertension in PISK was associated with lower deformation response, along with steepening and thickening of the cornea. [J Refract Surg. 2013;29(5):356–358.] From Instituto de Olhos Renato Ambrósio, Rio de Janeiro, Brazil (FF-C, IR, BV, AL, RA); Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Rio de Janeiro, Brazil (FF-C, IR, BV, AL, RA); Serviço de Oftalmologia, Hospital de São João, Porto, Portugal (FF-C); the Departments of Ophthalmology and Biomedical Engineering, The Ohio State University, Columbus, Ohio (CJR); and the Department of Ophthalmology, Federal University of São Paulo, São Paul, Brazil (RA). Drs. Ambrósio and Roberts are consultants for Oculus, Inc. The remaining authors have no financial or proprietary interest in the materials presented herein. AUTHOR CONTRIBUTIONS Study concept and design (FF-C, IR, BV, AL, RA, CJR); data collection (FF-C, IR, BV, AL, RA); analysis and interpretation of data (FF-C, IR, BV, AL, RA); drafting of the manuscript (FF-C); critical revision of the manuscript (IR, BV, AL, RA, CJR) Correspondence: Renato Ambrósio, Jr., MD, PhD, Rua Conde de Bonfim 211/712, Tijuca, Rio de Janeiro, RJ, Brazil 20520-050. E-mail: dr.renatoambrosio@gmail.com 10.3928/1081597X-20130129-03 To report the tomographic and biomechanical findings before and after treatment of a case of pressure-induced stromal keratopathy (PISK), which was misdiagnosed as diffuse lamellar keratitis (DLK). A case report of a referred patient with supposed diagnosis of DLK after LASIK in the right eye. Scheimpflug-based corneal tomography and biomechanical assessment were provided by the Pentacam HR and CorVis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany). A layer of corneal opacity beneath the flap with a presumably fluid-filled interface area was observed on slit-lamp biomicroscopy. Scheimpflug image from Pentacam revealed a hyperreflective area underneath the flap interface. Goldmann applanation tonometry was 12 mm Hg, whereas CorVis intraocular pressure was 53.5 mm Hg with deformation amplitude of 0.42 mm. Two days after starting oral and topical ocular hypotensive therapy, CorVis intraocular pressure was 14 mm Hg and deformation amplitude was 1.02 mm. Ocular hypertension in PISK was associated with lower deformation response, along with steepening and thickening of the cornea. [J Refract Surg. 2013;29(5):356–358.] From Instituto de Olhos Renato Ambrósio, Rio de Janeiro, Brazil (FF-C, IR, BV, AL, RA); Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Rio de Janeiro, Brazil (FF-C, IR, BV, AL, RA); Serviço de Oftalmologia, Hospital de São João, Porto, Portugal (FF-C); the Departments of Ophthalmology and Biomedical Engineering, The Ohio State University, Columbus, Ohio (CJR); and the Department of Ophthalmology, Federal University of São Paulo, São Paul, Brazil (RA). Drs. Ambrósio and Roberts are consultants for Oculus, Inc. The remaining authors have no financial or proprietary interest in the materials presented herein. AUTHOR CONTRIBUTIONS Study concept and design (FF-C, IR, BV, AL, RA, CJR); data collection (FF-C, IR, BV, AL, RA); analysis and interpretation of data (FF-C, IR, BV, AL, RA); drafting of the manuscript (FF-C); critical revision of the manuscript (IR, BV, AL, RA, CJR) Correspondence: Renato Ambrósio, Jr., MD, PhD, Rua Conde de Bonfim 211/712, Tijuca, Rio de Janeiro, RJ, Brazil 20520-050. E-mail: dr.renatoambrosio@gmail.com

This publication has 7 references indexed in Scilit: