Automatic selection of tidal volume, respiratory frequency and minute ventilation in intubated ICU patients as startup procedure for closed-loop controlled ventilation
Authors:
- 1 February 1994
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 11 (1) , 19-30
Abstract
Objective: Before a patient can be connected to a mechanical ventilator, the controls of the apparatus need to be set up appropriately. Today, this is done by the intensive care professional. With the advent of closed loop controlled mechanical ventilation, methods will be needed to select appropriate startup settings automatically. The objective of our study was to test such a computerized method which could eventually be used as a start-up procedure (first 5–10 minutes of ventilation) for closed-loop controlled ventilation. Design: Prospective Study. Settings: ICU's in two adult and one children's hospital. Patients: 25 critically ill adult patients (age≥15 y) and 17 critically ill children selected at random were studied. Interventions: To simulate ‘initial connection’, the patients were disconnected from their ventilator and transiently connected to a modified Hamilton AMADEUS ventilator for maximally one minute. During that time they were ventilated with a fixed and standardized breath pattern (Test Breaths) based on pressure controlled synchronized intermittent mandatory ventilation (PCSIMV). Measurements and main results: Measurements of airway flow, airway pressure and instantaneous CO2 concentration using a mainstream CO2 analyzer were made at the mouth during application of the Test-Breaths. Test-Breaths were analyzed in terms of tidal volume, expiratory time constant and series dead space. Using this data an initial ventilation pattern consisting of respiratory frequency and tidal volume was calculated. This ventilation pattern was compared to the one measured prior to the onset of the study using a two-tailed paired t-test. Additionally, it was compared to a conventional method for setting up ventilators. The computer-proposed ventilation pattern did not differ significantly from the actual pattern (p>0.05), while the conventional method did. However the scatter was large and in 6 cases deviations in the minute ventilation of more than 50% were observed. Conclusions: The analysis of standardized Test Breaths allows automatic determination of an initial ventilation pattern for intubated ICU patients. While this pattern does not seem to be superior to the one chosen by the conventional method, it is derived fully automatically and without need for manual patient data entry such as weight or height. This makes the method potentially useful as a startup procedure for closed-loop controlled ventilation.Keywords
This publication has 21 references indexed in Scilit:
- Pulmonary Function Indices in Critical Care Patients1988
- Closed-Loop Control of an Anesthesia Delivery System: Development and Animal TestingIEEE Transactions on Biomedical Engineering, 1987
- Computer-controlled optimization of positive end-expiratory pressureCritical Care Medicine, 1986
- A feedback controller for ventilatory therapyAnnals of Biomedical Engineering, 1985
- A Microcomputer-Based Differential Lung Ventilation SystemIEEE Transactions on Biomedical Engineering, 1982
- Systemic Arterial Blood pH Servocontrol of Mechanical VentilationAnesthesiology, 1978
- Computer control of respiration and anaesthesiaMedical & Biological Engineering & Computing, 1973
- An Optimally Controlled RespiratorIEEE Transactions on Biomedical Engineering, 1971
- CARBON DIOXIDE AND OXYGEN BLOOD LEVELS WITH A CARBON DIOXIDE CONTROLLED ARTIFICIAL RESPIRATORAnesthesiology, 1959
- CLINICAL USE OF A PHYSIOLOGICAL RESPIRATOR PRODUCING N2O AMNESIA-ANALGESIAAnesthesiology, 1957