Multi-modal surface registration for markerless initial patient setup in radiation therapy using microsoft's Kinect sensor
Authors:
- 1 November 2011
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1175-1181
Abstract
In radiation therapy, prior to each treatment fraction, the patient must be aligned to computed tomography (CT) data. Patient setup verification systems based on range imaging (RI) can accurately verify the patient position and adjust the treatment table at a fine scale, but require an initial manual setup using lasers and skin markers. We propose a novel markerless solution that enables a fully-automatic initial coarse patient setup. The table transformation that brings template and reference data in congruence is estimated from point correspondences based on matching local surface descriptors. Inherently, this point-based registration approach is capable of coping with gross initial misalignments and partial matching. Facing the challenge of multi-modal surface registration (RI/CT), we have adapted state-of-the-art descriptors to achieve invariance to mesh resolution and robustness to variations in topology. In a case study on real data from a low-cost RI device (Microsoft Kinect), the performance of different descriptors is evaluated on anthropomorphic phantoms. Furthermore, we have investigated the system's resilience to deformations for mono-modal RI/RI registration of data from healthy volunteers. Under gross initial misalignments, our method resulted in an average angular error of 1.5° and an average translational error of 13.4 mm in RI/CT registration. This coarse patient setup provides a feasible initialization for subsequent refinement with verification systems.Keywords
This publication has 22 references indexed in Scilit:
- Automatic Multi-modal ToF/CT Organ Surface Registration2011
- Characterization of a real‐time surface image‐guided stereotactic positioning systemMedical Physics, 2010
- Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancerRadiation Oncology, 2009
- Comparison of Target Registration Errors for Multiple Image-Guided Techniques in Accelerated Partial Breast IrradiationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Modeling respiratory motion variations in the 4D NCAT phantom2007
- Salient geometric features for partial shape matching and similarityACM Transactions on Graphics, 2006
- Distinctive Image Features from Scale-Invariant KeypointsInternational Journal of Computer Vision, 2004
- Efficient variants of the ICP algorithm2002
- Using spin images for efficient object recognition in cluttered 3D scenesIEEE Transactions on Pattern Analysis and Machine Intelligence, 1999