Selection of the optimum electrospray voltage for gradient elution LC-MS measurements
- 13 December 2008
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 20 (4) , 682-688
Abstract
Changes in liquid composition during gradient elution liquid chromatography (LC) coupled to mass spectrometry (MS) analyses affect the electrospray operation. To establish methodologies for judicious selection of the electrospray voltage, we monitored in real time the effect of the LC gradient on the spray current. The optimum range of the electrospray voltage decreased as the concentration of organic solvent in the eluent increased during reversed-phase LC analyses. These results and related observations provided the means to rationally select the voltage to ensure effective electrospray operation throughout gradient-elution LC separations. For analyses in which the electrospray was operated at constant voltage, a small run-to-run variation in the spray current was observed, indicating a changing electric field resulting from fouling or degradation of the emitter. Algorithms using feedback from spray current measurements that can maintain the electrospray voltage within the optimum operating range throughout gradient elution LC-MS were evaluated. The electrospray operation with voltage regulation and at a constant, judiciously selected voltage during gradient elution LC-MS measurements produced data with similar reproducibility.Keywords
This publication has 22 references indexed in Scilit:
- Decon2LS: An open-source software package for automated processing and visualization of high resolution mass spectrometry dataBMC Bioinformatics, 2009
- Analytical Characterization of the Electrospray Ion Source in the Nanoflow RegimeAnalytical Chemistry, 2008
- Nanoelectrospray Emitter Arrays Providing Interemitter Electric Field UniformityAnalytical Chemistry, 2008
- Fully Automated Four-Column Capillary LC−MS System for Maximizing Throughput in Proteomic AnalysesAnalytical Chemistry, 2007
- Electrospray Characteristic Curves: In Pursuit of Improved Performance in the Nanoflow RegimeAnalytical Chemistry, 2007
- Oxidation Artifacts in the Electrospray Mass Spectrometry of Aβ PeptideAnalytical Chemistry, 2007
- Chemically Etched Open Tubular and Monolithic Emitters for Nanoelectrospray Ionization Mass SpectrometryAnalytical Chemistry, 2006
- Robust Algorithm for Alignment of Liquid Chromatography−Mass Spectrometry Analyses in an Accurate Mass and Time Tag Data Analysis PipelineAnalytical Chemistry, 2006
- Mass spectrometry-based proteomicsNature, 2003
- An Electrohydrodynamic Ion Source for the Mass Spectrometry of LiquidsReview of Scientific Instruments, 1972