Multisource inverse-geometry CT. Part I. System concept and development
Authors:
- 18 July 2016
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 43 (8Part1) , 4607-4616
Abstract
Purpose: This paper presents an overview of multisource inverse-geometry computed tomography (IGCT) as well as the development of a gantry-based research prototype system. The development of the distributed x-ray source is covered in a companion paper [V. B. Neculaes et al., “Multisource inverse-geometry CT. Part II. X-ray source design and prototype,” Med. Phys. 43, 4617–4627 (2016)]. While progress updates of this development have been presented at conferences and in journal papers, this paper is the first comprehensive overview of the multisource inverse-geometry CT concept and prototype. The authors also provide a review of all previous IGCT related publications. Methods: The authors designed and implemented a gantry-based 32-source IGCT scanner with 22 cm field-of-view, 16 cm z-coverage, 1 s rotation time, 1.09 × 1.024 mm detector cell size, as low as 0.4 × 0.8 mm focal spot size and 80–140 kVp x-ray source voltage. The system is built using commercially available CT components and a custom made distributed x-ray source. The authors developed dedicated controls, calibrations, and reconstruction algorithms and evaluated the system performance using phantoms and small animals. Results: The authors performed IGCT system experiments and demonstrated tube current up to 125 mA with up to 32 focal spots. The authors measured a spatial resolution of 13 lp/cm at 5% cutoff. The scatter-to-primary ratio is estimated 62% for a 32 cm water phantom at 140 kVp. The authors scanned several phantoms and small animals. The initial images have relatively high noise due to the low x-ray flux levels but minimal artifacts. Conclusions: IGCT has unique benefits in terms of dose-efficiency and cone-beam artifacts, but comes with challenges in terms of scattered radiation and x-ray flux limits. To the authors’ knowledge, their prototype is the first gantry-based IGCT scanner. The authors summarized the design and implementation of the scanner and the authors presented results with phantoms and small animals.Keywords
Funding Information
- General Electric (Internal)
- National Institute of Biomedical Imaging and Bioengineering (EB006837S, R01 EB006837)
- General Electric (GE) (Internal)
- National Institute of Biomedical Imaging and Bioengineering (NIBIB) (EB006837S, R01 EB006837)
- General Electric (Internal)
- National Institute of Biomedical Imaging and Bioengineering (EB006837S, R01 EB006837)
This publication has 30 references indexed in Scilit:
- A new reconstruction method to improve SNR for an inverse geometry CT system2008
- 3D iterative full and half scan reconstruction in CT architectures with distributed sources2008
- Iterative reconstruction for multi-source inverse geometry CT: a feasibility study2007
- Analytical cone-beam reconstruction using a multi-source inverse geometry CT system2007
- A prototype table-top inverse-geometry volumetric CT systemMedical Physics, 2006
- First performance evaluation of a dual-source CT (DSCT) systemEuropean Radiology, 2005
- A three‐dimensional reconstruction algorithm for an inverse‐geometry volumetric CT systemMedical Physics, 2005
- An inverse‐geometry volumetric CT system with a large‐area scanned source: A feasibility studyMedical Physics, 2004
- Scanning-beam digital x-ray (SBDX) system for cardiac angiography1999
- Large-area scanning x-ray source to maximize contrast sensitivity by minimizing scatter detection1993