Randomized Controlled Trial of Perioperative ICD Management: Magnet Application versus Reprogramming
Authors:
- 7 May 2014
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 37 (9) , 1219-1224
Abstract
Background: There are insufficient data to guide perioperative implantable cardioverter‐defibrillator (ICD) management for patients undergoing surgical procedures using electrocautery.Methods: We conducted a multicenter randomized controlled trial of patients with ICDs undergoing surgery with monopolar electrocautery. Subjects were randomized to an “Off” group (ICD therapy programmed off, then postoperatively programmed on) or a “Magnet” group (ICD therapy suspended with a magnet and no immediate postoperative ICD interrogation). Also, a registry was maintained of ICD patients with procedures within 6 inches of the ICD (all programmed off). The primary endpoint was ICD off time with secondary endpoints being caregiver handoffs and incidence of electromagnetic interference (EMI).Results: All patients (n = 80) had pectoral ICDs. Subject demographics were well matched in each group, and duration of electrocautery was similar (80 minutes vs 64 minutes, P = 0.58). The mean “excess” ICD off time (ICD off time – electrocautery time) was significantly higher in the Off group than the Magnet group (115 minutes vs 28 minutes, P < 0.001). Mean number of caregiver handoffs were higher in the Off group (6.6 vs 5.5, P < 0.001). There was no EMI in any lower abdominal or lower extremity procedures. Neither group had arrhythmic events or device reset.Conclusion: A magnet protocol simplifies perioperative ICD management for procedures using electrocautery more than 6 inches from the ICD. This protocol results in significantly shorter time with ICD therapy off, fewer provider handoffs, no risk of inadvertently discharging patients home with ICD therapies off, and no device reset.Keywords
Funding Information
- Medtronic
- St. Jude Medical
This publication has 9 references indexed in Scilit:
- The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the Perioperative Management of Patients with Implantable Defibrillators, Pacemakers and Arrhythmia Monitors: Facilities and Patient ManagementHeart Rhythm, 2011
- A Systematic Review of Failures in Handoff Communication During Intrahospital TransfersThe Joint Commission Journal on Quality and Patient Safety, 2011
- The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) TrialJACC, 2011
- Hand‐Off Communication: A Requisite for Perioperative Patient SafetyAORN Journal, 2008
- Effects of Surgical and Endoscopic Electrocautery on Modern‐Day Permanent Pacemaker and Implantable Cardioverter‐Defibrillator SystemsPacing and Clinical Electrophysiology, 2008
- Electromagnetic Interference (EMI) and Arrhythmic Events in ICD Patients Undergoing Gastrointestinal ProceduresPacing and Clinical Electrophysiology, 2007
- Deaths associated with implantable cardioverter defibrillator failure and deactivation reported in the United States Food and Drug Administration Manufacturer and User Facility Device Experience DatabaseHeart Rhythm, 2004
- Application of Radiofrequency Energy in Surgical and Interventional Procedures:Pacing and Clinical Electrophysiology, 2004
- Interference in Implanted Cardiac Devices, Part IIPacing and Clinical Electrophysiology, 2002